background image

Data logger operating manual

7

EN

Model TCO

The TCO data logger has three internal sensors to detect measured values from
a total of seven channel groups. These channel groups (measured values) are:
Air temperature in °C, air temperature in °F, dew point in °C, dew point in °F, rel-
ative humidity in %, absolute humidity in g/m³ and CO

2

concentration in ppm.

For  each  channel  group,  there  are  four  measuring  channels  for  recording:
Current measured value (cur), minimum measured value (min), maximum meas-
ured value (max) and average measured value (mid). In total, there are 28 meas-
uring channels available for your TCO data logger, as shown in table 1.

7.2.    Displaying measured values

One of the channel groups specified in table 1 for displaying measured values
can be configured to be displayed on each of the three measured value rows.
Here, the current measured value is always shown on the display.

7.3.    Data recording

If either of the operating modes M3 or M4 are selected, then the data logger is
in logging mode (REC) and the measured values of the measuring channels se-
lected for recording are saved in the device. 

Up to 20 of the measuring channels specified in table 1 can be simultaneously
recorded in the data memory of the measuring device. 

Recording begins immediately from the moment that the operating mode M3 or
M4  is  selected  and  is  carried  out  in  ring  mode.  This  means  that  when  the
memory limits are reached, the recording does not stop but continues to record.
The older values are simply overwritten by the newest measured values.

Tip: Current, minimum, maximum and average measured values for

a channel group each represent one measuring channel. If these

values should be available for later documentation and evaluation, each of
the measuring channels must already be selected during memory organi-
sation for recording, because it is not possible to calculate these values 
in the software.

The specifications for type, duration and scope of data recording in the logging
mode can be individually set in the software. Detailed information is provided in
the software manual.

8. Alarm function

8.1.    Alarm configuration

An alarm function can be configured in the measuring device administration for
each of the measuring channels selected for display or recording. 

By defining an upper and lower alarm limit value, a value corridor is specified.
This is the so-called good range and when the range is exited, an alarm sounds. 

Note: The alarm function can only be represented on the display for the chan-
nel  groups  of  the  three  measuring  channels  selected  for  measured  value
 display, and only in the operating modes M2 and M3! 

However, saving alarm results of other selected measuring channels in logging
mode  is  independent  of  the  selected  display  of  the  measured  values.  That
means  that  the  alarm  results  of  the  measuring  channels  which  are  not  for
display but for data recording are also recorded. 

Even in the operating mode M4 when displaying measured values is deactivated,
all preselected alarm results are recorded!

8.2.    Alarm display

The  alarm  display  is  only  possible  for
measured values which are preselected
for displaying and which have been con-
figured  with  an  active  alarm  function.

If an alarm occurs at one of these meas-
uring  channels,  the  alarm  symbol
flashes beside the measured value dis-
play  for  this  channel  while  the  good
range is exited. 

If the measured value then reaches the

defined  value  corridor,  that  is,  the  good  range,  then  the  alarm  symbol  stops
flashing and is shown continually. 

This  indicates  that  an  alarm  result  has  occurred.  If  the  good  range  is  exited
during the course of a measurement, the alarm symbol flashes again.

Reading the measured value memory deletes the alarm display.

Additionally, a hysteresis can be set by which the measured value must return
to the valid range to switch off the alarm.

8.3.    Acoustic alarm

If an alarm goes off while the acoustic function is activated (see chapter 6.2.4.
Acoustic function), then an alarm tone sounds and only stops when the measured
value returns to the good range.

Example display for an alarm result of the

measuring channel shown in the first

measured value row.

Alarm display

Upper alarm 
limit value

Valid value
 corridor 
“good range”

Lower alarm
limit value

flashing

static

Logging mode

Reading the measured value memory

Содержание TCO

Страница 1: ...BA THI THIP TCO 05 EN v1 0 Modification rights reserved Turfschipper 114 2292 JB Wateringen Tel 31 0 174 272330 www catec nl info catec nl Turfschipper 114 2292 JB Wateringen Tel 31 0 174 272330 www...

Страница 2: ...table recording or re quest intervals then saved and transmitted to a connected PC The measuring device may only be used for this intended use while complying with specified technical data Any other u...

Страница 3: ...Display symbol for power supply over a network Display symbol for power supply over USB Display for battery level Display for mode marker Display for active measured value recording Display for inacti...

Страница 4: ...y state the power consumption is minimal because measured values cannot be requested or shown M53 M52 M51 M4 M3 M2 M1 Selection of the operating mode M1 Display of the operating mode M1 5 Preparation...

Страница 5: ...on is ac tivated then software configuration and data reading from the data logger can be carried out over the network A signal which is broadcast from the data logger over UDP allows the SmartGraph s...

Страница 6: ...minimum measured value min maximum meas ured value max and average measured value mid In total there are 24 meas uring channels available for your THI data logger as shown in table 1 Model THIP The TH...

Страница 7: ...ed in the software manual 8 Alarm function 8 1 Alarm configuration An alarm function can be configured in the measuring device administration for each of the measuring channels selected for display or...

Страница 8: ...ed batteries in the household rubbish or throw them in the fire instead dispose of them according to the relevant legal requirements 9 2 Deleting measured data The detected measured data are permanent...

Страница 9: ...r at 25 C 0 5 hPa Screen resolution 0 1 hPa CO2 concentration Operating principle NDIR Measuring range 0 5 000 ppm Accuracy 50 ppm 3 measured value at 20 C and 1 013 mbar Screen resolution 1 ppm Long...

Отзывы: