Starting up your SONUS Dosemeter 

To turn on your instrument press and hold the Power On/Off key 

 for approximately one 

second.  (Your meter will display the start-up screen and initialise any saved settings.) 

 

To  Calibrate  using  a  Castle  Calibrator  (GA601  or  GA607)  attach  the  calibrator  to  the 
microphone, press the 

MENU

 key and use the 

 



arrow keys as required until  CAL is on 

the  screen.  Press 

OK

  twice  and  the  instrument  will  calibrate  to  the  level  of  the  calibrator 

(default 94dB), displaying 

COMPLETE

 when finished. 

 

Setting the Time-history interval 

 

Press 

MENU

 then press the 



arrow key to find 

SETUP

. Press 

OK

 to show 

INTERVAL

 on 

screen, then press 

OK

 again. Using the 

 



arrow key, select the desired time-history size 

and press 

OK

. Display will show

 DONE

. Press 

C

 to return to measuring screen. 

 

Taking a Measurement 

To  start  recording,  press  the 

STOP/RECORD

  Key 

  so  the  display  briefly  shows 

REC

 

MODE

. You can review the parameters using 

 



arrow keys during recording 

When ready to stop, press 

 again, so the display shows 

SAVE YES

. If you don’t want to 

save, press 

C

 key. To save, press 

OK

 

The following will then be displayed: - 

 

Display 

Description 

8Hr ET 

Set Exposure Time period to 8 Hours 

ENTER ET 

Enter an Exposure Time period 

NO ET 

Use Measurement Run Time period (Log Length) 

 

Exposure Time entry uses the following 24 hour format (hh:mm:ss) and the adjustable figure 
flashes, this always starts with the 

HOUR

 figure. 

 

Locking the Keypad 

To lock the keypad, press the padlock key 

. This will immediately lock the keypad, and the 

display will show the battery indicator screen, i.e. BAT: |||| |||| |||| ||||. To unlock the instrument, press 

C, OK, i

 in succession, and the display will return to the LAF screen. 

 

Reviewing measurements 

 

Press the 

MENU

 key then use the 

 

 

arrow keys until screen displays 

‘LOGS’

 (If no logs are 

available this option is not shown). 
Press 

OK

 and the instrument will display the currently saved logs in date order. 

Scroll through the available dates using the 

 

 

arrow keys. 

Press 

C

  to  exit  without  selection  or 

OK

  to  confirm  the  selection,  the  instrument  will  then 

display the currently saved logs by time recorded. 
Scroll through the available times using the 

 

 

arrow keys. 

Press 

C

 to exit without selection and return to the date list or 

OK

 to confirm the selection, 

the instrument will then view the saved cumulative information for the log selected. 
Scroll through the available data using the 

 

 

arrow keys. 

Press 

C

 to exit and return to the parameter display screen when required.

 