Dust Detective             

                                                                                                

 

 

 
 

 

10 

 

Starting measurements 

Checking the Instrument’s Zero

 

Check the zero calibration for the Microdust Pro before every use. This can be achieved either as described on 
Page 28 of the 

Microdust Pro

 

Handbook

, utilising the purge air bellows connected to the probe purge point. 

Alternatively, to zero in situ, remove the inlet cowl, and connect the clean air filter to the inlet tube, start the 
Apex2 pump. With clean air in the Probe, the display should be zeroed. 

Checking Probe Span 

The span of the Microdust Pro should be checked in accordance with Page 31 of the 

Microdust Pro Handbook

The span can also be checked by inserting the calibration filter and checking the reading. If required, adjust 
the span control according to the handbook. 

Sampling Pump 

The Dust Detective enclosure is designed to use the Apex2 range of personal samplers. This allows the user to 
take a standard personal sampling pump, charge the unit up and install inside the enclosure. The pumps are 
designed to provide a flow rate capability of up to 5 L/min whilst using a 37 mm GFA filter. Automatic flow 
control circuitry is used to maintain a stable sample flow rate over varying pressure drop conditions. PUF 
filters are designed to operate at a flow rate of up to 3.5 L/min, refer to the 

Apex 2 Handbook

 (page 13) for 

details on changing the flow rate. 

To switch the pump on, press the power key on the pump. To start the pump running press the Enter Key. 
When running, flow rate will appear on the display and a blue flashing light will be visible at the top of the 
pump. To stop the pump, hold the Enter key again. 

To switch the pump off, press and hold the Power key and the pump will countdown to off.  

If the pump is unable to maintain the selected flow rate due to excessive pressure drop or an inlet blockage, 
the red LED will begin to flash and the warning tone will bleep rapidly.  

Starting a Sample Run 

1.

 

Install the sample inlet tube/head to the case (see Figure 1). 

2.

 

Insert the PUF filter foam(s) and in line sample filter (if required) (see Figure 4 & 5). 

3.

 

Switch the Microdust and Apex2 on. Confirm the Zero and Span settings are correct as described 
above. 

4.

 

Set the Apex2 flow rate to 2L/min 

5.

 

With the unit located for measurement, start the Microdust logging by holding the Play key and start 
the pump by holding down the Enter key. 

6.

 

Shut and lock the Dust Detective Case. 

 

Содержание Dust-Detective

Страница 1: ...L Static Air Sampling Enclosure HB3341 07 November 2018 Casella Regent House Wolseley Road Kempston Bedford MK42 7JY T 44 0 1234 844 100 F 44 0 1234 841 490 E info casellasolutions com www casellasolu...

Страница 2: ...ation 7 Storage of PUF Filters 7 PUF Filter Adaptor 8 Inserting Removing PUF Filters 8 Collection Filter Cassette 9 Starting measurements 10 Checking Probe Span 10 Sampling Pump 10 Service 11 Technica...

Страница 3: ...d of by placing in a fire or incinerator nor must they be punctured crushed or otherwise mutilated or opened up in any way Disposal must be in accordance to local environmental legislation Ensure the...

Страница 4: ...from those in the manual If there are any questions please contact Casella for clarification Casella makes continual advancements in its products and services We therefore reserve the right to make ch...

Страница 5: ...ssory provides a simple solution for short to medium term fixed area monitoring with the Microdust Pro and Apex2 sampling pumps and is designed specifically for use in indoor applications but some sho...

Страница 6: ...cutive publication MDHS 14 4 The foam filter specifications and dimensions determine the desired aerosol size selection characteristics and eliminate particle sizes greater than PM10 PM2 5 or Respirab...

Страница 7: ...m producing spurious results The system incorporates an Apex2 sampling pump to draw the sample air through the inlet pipe at a user selected flow rate PUF filters are designed to operate at a flow rat...

Страница 8: ...ise damage to the instruments could occur Connect the Apex2 pump to the sampling spigot as illustrated in Figure 1 Connect the Microdust probe into the Adaptor Figure 5 When locked correctly into posi...

Страница 9: ...a specified mean aerodynamic size it is also possible to weigh the PUF inserts before and after measurement to establish both the total suspended particulate TSP value and the desired size fraction In...

Страница 10: ...e Always handle filters with care to prevent contamination or physical damage It is recommended that even if gravimetric analysis is not being performed a GFA filter P102051 is still used to prevent p...

Страница 11: ...atic flow control circuitry is used to maintain a stable sample flow rate over varying pressure drop conditions PUF filters are designed to operate at a flow rate of up to 3 5 L min refer to the Apex...

Страница 12: ...d premises We will be happy to Provide quotations for individual repairs or provide annual maintenance under contract We recommend factory service by technicians trained and equipped to repair your in...

Страница 13: ...ge humidity range 0 to 90 RH in the absence of condensation Size 410 x 330 x 175 mm Apex2 Flow range 0 8 to 5 0 L min Flow control accuracy 5 for selected flow Status indicators Red Blue LED Dimension...

Страница 14: ...Spare clear cassette P102113 Filter support pad 37mm pack of 100 P102051 Gravimetric GFA filter 37mm pack of 100 TUC139 Spigot 90 Degree from clear cassette TUC133 Luer connector from TUC133 to ID tu...

Страница 15: ...casellasolutions com WEEE INFORMATION FOR EU MEMBER STATES ONLY The use of the WEEE symbol indicates that this product may not be treated as household waste By ensuring this product is disposed of cor...

Отзывы: