CASELLA CEL CEL-62 Series Скачать руководство пользователя страница 4

 

Casella CEL

 

1.

 

INTRODUCTION 

 

CEL-62X Series is a range of Sound Level Meters designed to meet the demands of 
worldwide industrial health and safety professionals as well as general noise 
measurement applications. The instrument encompasses functionality from basic sound 
level measurement to integrating and real-time octave band analysis. 

The CEL-62X instrument is based on the very latest digital signal processing technology 
and has a crisp, vibrant colour TFT LCD display. This precision instrument is designed 
to bring your noise measurements up to date giving confident and reliable performance 
fully compliant with international standards. 

Data is stored in a .csv format, compliant with MS Office applications, and through the 
USB connection, the .csv files can be copied to a PC without the need for any propriety 
software packages. 

Back to 

Index

2. INSTRUMENT FEATURES 

 

The CEL-62X Sound Level Meter comprises of several variants, where the ‘X’ 
represents a number which determines the model variant and subsequent features. 
Subsequent letters denote the frequency analysis capability, ‘A’ for broadband, ‘B’ for 
1/1 octave band and ‘C’ for 1/3 octave, e.g. CEL-620C. 

CEL-620 

An integrating meter providing A, C, and Z frequency weightings, Fast Slow and Impulse 
time weightings with cumulative storage and PC connectivity functions. A single 
measurement range of 140dB RMS is standard. Simultaneous L

Aeq

 and L

Ceq

 values are 

measured for use in the HML selection of hearing protection method. L

avg

 is available 

with a selectable threshold. This model is available in Class 1 or Class 2 accuracy 
standards. 

CEL-621 

Additional parameters are added for the measurement of environmental noise. 

 

Statistical parameters (L

n%

) are available and can be altered to suit specific 

environmental legislation. 

Frequency Analysis

 

Models providing ‘Integrating’ functionality combined with addition of 1/1 Octave bands 
(B model) or both 1/1 Octave and 1/3 Octave bands (C model) analysis. This instrument 
is designed to satisfy the demands of all industrial noise at work applications. Octave 
results may be displayed graphically or in numerical form and can be A, C or Z 
frequency weighted. Advanced DSP processing provides fast parallel processing of 
Octave results from 16Hz to 16kHz, or 1/3 Octave results from 12.5Hz to 20kHz. This 
model is available in Class 1 or Class 2. 

 
DESCRIPTION (Refer to Figure 1) 

 

The CEL-62X has a compact, ergonomic and robust ‘rubber grip’ case design that sits 
the instrument comfortably in the hand when not mounted on a tripod. The ½” 
microphone (under windshield 1) can be removed from the fixed preamplifier (2). The 
windshield (1) should be used to cover and protect the microphone at all times.  
When the instrument is switched ON (3) it will run through its own software initialisation 
and then default to a Stop mode (Red Bar) screen. 

 

Содержание CEL-62 Series

Страница 1: ...Eur polis Calle C n 4B 28230 Las Rozas Madrid Spain Phone 34 91 640 75 19 Fax 34 91 636 01 96 E mail online casella es com Web www casella es com April 2010 Web www casellameasurement com 17 Old Nash...

Страница 2: ...Casella CEL 2...

Страница 3: ...2 3 F Set Run Duration 14 3 G Calibration ref Level 15 3 H View Memory Results 16 3 I Delete Memory results 17 3 J Run Screens 18 3 K Instrument Calibration 19 3 L Extracting Data 20 3 M System Tools...

Страница 4: ...vity functions A single measurement range of 140dB RMS is standard Simultaneous LAeq and LCeq values are measured for use in the HML selection of hearing protection method Lavg is available with a sel...

Страница 5: ...ation The CEL 62X can be explored just like a USB memory device and the data files of the individual measurements are saved in a CSV Comma Separated Variable format that is compatible with MS Office a...

Страница 6: ...e automatically used instead of the batteries When downloading data to a PC the CEL 62X can be powered by the USB connection The 12VDC or USB connection will not re charge batteries With 12VDC or USB...

Страница 7: ...sion instrument and should be calibrated before use to ensure accurate measurements Also care should be taken that the operator does not unduly affect the sound field Ideally the instrument should be...

Страница 8: ...models have both 1 1 octave and 1 3 octave functionality In order to change between 1 1 octave and 1 3 octave use the Cursor Keys C to navigate to the Octave icon 3 and press Select B Use the Cursor K...

Страница 9: ...screen you can define the following RESPONSE Random Freefield Random microphone response is typically used to conform to US legislation for workplace noise whereas Freefield is used throughout the EU...

Страница 10: ...To select the Main value press and hold the Run Stop Key D until the parameter is highlighted in red Other parameters can be added Navigate to the appropriate parameter that is to be added A rectangl...

Страница 11: ...wn Cursor Keys C to adjust the settings and the Left Right Cursor Keys C to move between the hours minutes and seconds When the correct time has been set press Save B SET DATE Press the Up Down Cursor...

Страница 12: ...HT From the Menu screen use the Cursor Keys C to navigate to the Settings icon 1 Use the Cursor Keys C to navigate to the Backlight icon 2 Press Select B Figure 7 Set Backlight Sheet 1 ACTIVATION The...

Страница 13: ...the Duration has been set press Save B Figure 7 Set Backlight Sheet 2 BRIGHTNESS The backlight brightness affects the battery life For improved battery performance use the lowest brightness level sui...

Страница 14: ...ield highlighted Press Edit B to set the Duration timer Use the Left Right Cursors C to toggle between either On or Off Press Save B to continue Figure 8 Run Duration Sheet 1 DURATION Press the Up Dow...

Страница 15: ...om 93 95dB or 113 115dB A 1kHz calibrator must be used From the Menu screen use the Cursor Keys C to navigate to the Settings icon 1 Use the Cursor Keys C to navigate to the Cal Ref Level icon 2 Press...

Страница 16: ...verview of measurement data for the selected a Run This is read only To see the specific results press View B Figure 10 View Memory Results Sheet 1 The Octave values CEL 62XB and C models only saved i...

Страница 17: ...rsor Keys C to navigate to the Delete icon 2 and press Select B This screen asks the operator to either delete only the current run or to delete all the runs within the memory Use the Up Down Cursors...

Страница 18: ...ft hand corner of the screen The left hand Soft Key A is the Pause Run feature When Pause is selected Paused will be displayed on the screen The incremental run number and duration are displayed in th...

Страница 19: ...5 Note The Instrument must be at a STOP screen before the Acoustic Calibrator is fitted For calibrator reference level adjustment refer to Section G Calibrations should be performed before and after m...

Страница 20: ...these files regardless of the displayed parameters selected within the instrument setup The CEL 62X acts as a removable storage device when connected to a PC by the CMC51 cable See Figure 18 for the...

Страница 21: ...t Note that this is primarily used for testing purposes Format All Memory will format the USB memory CAUTION All measurement data will be lost Use the Cursor Keys C to highlight the required option an...

Страница 22: ...0 10 Lsg 10 Where Lact Actual corrected sound level Lms Measured sound level Lsg Self generated noise level High Level Sound Overload When a sound level is detected that is above the linear operating...

Страница 23: ...6 software PEAK MEASUREMENT A C and Z weighted from 65 0 to 143 3dB RMS DETECTOR Digitally derived True root mean square detection RMS 0 1dB display resolution NOISE FLOOR Total inherent Noise includi...

Страница 24: ...hours in Broadband mode with backlight continuously on low intensity High backlight setting will have a detrimental affect on battery life 12 hours measuring with backlight off INTERNAL CLOCK Date an...

Страница 25: ...onds to the Z weighted response of the sound level meter DC OUTPUT Via 2 5mm stereo audio jack Barrel ground Ring DC output provided for chart recorders loggers etc note some offset and scaling will b...

Страница 26: ...LEQ LWEQ LAEQ NA NA NA SPL value changes to LEQ during a measurement run Simultaneous C A integrating L w Pk LCPEAK NA NA NA A peak only available on user setups Lpeak LIEQ integrating L w IEQ LAIEQ N...

Страница 27: ...LAF Parameters LAeq Lavg Lavg LAeq T80 LAeq Lavg LAeq LAeq LCeq LASmax LASmax LASmax LCeq LCeq LAFmax LAFmax LC Aeq LZpeak LZpeak LZpeak LC Aeq LASmax LZpeak LZpeak LAFmax LC Aeq LC Aeq LC Aeq LAFmax...

Страница 28: ...ese undertakings will be passed on to the user CASELLA CEL s liability is limited to items of their own manufacture and they do not accept liability for any loss resulting from the operation or interp...

Страница 29: ...ectively giving an average level over the measurement period Following the ISO procedures doubling the energy results in a 3dB change in the Leq This is denoted by exchange rate Q 3 For example if the...

Страница 30: ...ue would be calculated to be 87dB A as the measurement duration is half the 8 hour Criterion Time and the exchange rate is 3 dB LCeq The C weighted equivalent level is the level that would contain the...

Страница 31: ...data for the CEL 251 microphone used for periodic testing Frequency 0o Free Field Corrections using a Bruel Kjear 4226 Calibrator 0o Free Field Corrections with Wind Screen using a Bruel Kjear 4226 Ca...

Страница 32: ...rical Typical dB Electrical Max dB Microphone Thermal dB Combined Typical dB Combined Max dB A 18 0 20 0 16 0 20 0 22 0 C 21 0 23 0 16 8 22 0 24 0 Z 28 0 30 0 16 8 28 0 30 5 Note The CEL 252 microphon...

Страница 33: ...15 seconds after a 5kPa change in ambient pressure m Electric Field Strength Greater Than 10 V m The CEL 62X was not tested for field strengths greater than 10 V m n EMC Emissions Emissions were no gr...

Страница 34: ...8 0 2 0 2 0 5 0 7 0 7 0 22 3150 3162 27 0 3 0 3 0 6 0 3 0 3 0 22 3550 3548 13 0 9 0 9 0 7 0 2 0 2 0 23 4000 3981 07 0 0 0 0 0 7 0 7 0 7 0 23 4500 4466 83 0 0 0 0 0 7 0 7 0 7 0 25 5000 5011 86 0 1 0 1...

Страница 35: ...20 0 48 0 55 0 22 0 66 4000 3981 07 0 00 0 10 0 58 1 00 0 75 0 96 0 29 1 17 1 17 1 84 1 75 1 67 1 22 1 86 1 32 1 90 4500 4466 83 0 00 0 12 0 26 0 18 0 57 1 11 0 76 1 06 1 06 1 70 2 28 2 28 1 23 1 70 1...

Страница 36: ...53 0 38 0 31 0 26 0 45 0 49 0 62 0 59 4000 3981 07 0 00 0 27 0 82 1 02 0 98 0 98 0 18 0 16 1 03 1 78 1 55 1 66 1 13 1 64 0 51 1 49 4500 4466 83 0 00 0 14 0 10 0 57 0 96 2 12 1 25 0 69 1 10 2 06 2 95 2...

Страница 37: ...Type 1 approved Cirrus CR 515 0 1dB 0 0dB Type 1 approved Examples The CEL 62X can be calibrated at a nominal level of 94dB or 114dB dependent on the type of calibrator used If a CEL 110 1 is used an...

Отзывы: