S
HELVING
I
NSTALLATION
Heating in a vacuum environment takes place primarily through conduction. Heat is transported from
the heating elements to the samples or product through the shelves. Install the shelves as described
below to ensure proper heat conduction and temperature measurement.
Never place samples or product on the oven chamber floor. The floor runs hotter than the shelf
temperatures. All heating specifications for the oven are for shelving temperatures.
CVO-2-HT
1.
Carefully slide the short shelf into position on the chamber floor, sliding the clip on the bottom
of the shelf onto the oven temperature probe.
•
The oven probe extends from the back wall near the floor of the chamber.
•
The short shelf must be on the bottom
of the shelf-stack to ensure the oven meets its
temperature uniformity specifications.
2.
Place the 2 tall shelves on top of the short shelf.
Continued on next page
Temperature Probe Shelf Clip
Bottom of Short Shelf
Содержание CVO-2-HT
Страница 1: ...Vacuum Ovens CVO 2 HT CVO 5 HT Installation Operation Manual...
Страница 2: ...2 P a g e cascadesciences com Tel 503 847 9047 Pictured on front cover left to right CVO 2 HT CVO 5 HT...
Страница 6: ...6 P a g e cascadesciences com Tel 503 847 9047 TABLE OF CONTENTS...
Страница 8: ...8 P a g e cascadesciences com Tel 503 847 9047 CERTIFICATIONS...
Страница 12: ...12 P a g e cascadesciences com Tel 503 847 9047 UNIT SPECIFICATIONS...
Страница 16: ...16 P a g e cascadesciences com Tel 503 847 9047 INTRODUCTION...
Страница 22: ...22 P a g e cascadesciences com Tel 503 847 9047 RECEIVING YOUR OVEN...
Страница 32: ...32 P a g e cascadesciences com Tel 503 847 9047 GRAPHIC SYMBOLS...
Страница 71: ...71 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS...
Страница 72: ......