![Casabeds Casa Med Classic 4 Скачать руководство пользователя страница 32](http://html1.mh-extra.com/html/casabeds/casa-med-classic-4/casa-med-classic-4_operating-manual_2566247032.webp)
Additional_Information_EMC_IPROXXX_EN60601-1-2_REV00.docx
Page1 of 7
A d d i t i o n a l I n f o r m a t i o n
–
Electromagnetic compatibility
–
(*IEC/EN 60601-1-2)
Date: February 27, 2013
REV.: 00
Made by: W. Kracht
Additional information intended to be used by the end-product manufacturer
(See *IEC/EN 60601-1-2)
Additional information intended to be used with DEWERT actuator system consists of DEWERT
component(s)/product(s):
•
Handset IPROXX
The above described component(s)/product(s) is/are hereinafter called
PRODUCT
.
NOTE: These instructions are only intended to be used by the end-product manufacturer.
They should not be given to the operator of the end product. The factual information
contained within may be used as a basis when creating the end-product manual.
Observe the document “Installation Instruction“ of PRODUCT (see above).
* Standard(s):
•
IEC 60601-1-2:2007
•
EN 60601-1-2:2007
The standards (see above) hereinafter called IEC/EN 60601-1-2
A d d i t i o n a l I n f o r m a t i o n
–
Electromagnetic compatibility
–
(*IEC/EN 60601-1-2)
Additional information intended to be used by the end-product manufacturer
(See *IEC/EN 60601-1-2)
Additional information intended to be used with DEWERT actuator system consists of DEWERT product(s),
such as:
%
Control Unit, e.g. MC10, MC11, MCL, MCL II
%
Single Drive, e.g. MEGAMAT MCZ, DS10
The above described DEWERT system consists of DEWERT product(s) is/are hereinafter called
PRODUCT
.
* Standard(s):
%
IEC 60601-1-2:2007
%
EN 60601-1-2:2007
The standards (see above) hereinafter called IEC/EN 60601-1-2
1
Guidance and manufacturer‘s declaration
– Electromagnetic Emissions -
(*IEC/EN 60601-1-2, Table 1)
* edition: see page 1 – standard(s)
2 The PRODUCT is intended for use in the electromagnetic environment specified below. The customer
or the user of the PRODUCT should assure that it is used in such an environment.
3
Emissions test
4 RF emissions
CISPR 11
Group 1
The PRODUCT uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
6 RF emissions
CISPR 11
Class B
7 Harmonic emissions
IEC 61000-3-2
Class A
8 Voltage fluctuations/ flicker
emissions
IEC 61000-3-3
Complies
The PRODUCT is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public
lowvoltage power supply network that supplies
buildings used for domestic purposes.
0B0$43;0BB82*0B0$43;0BB82#>F0B0$43,;CA0*M
6I+-
43656 - Patterson Medical 22069 Booklet 11-11.indd 26
11/11/11 11:40:01
Casa Med Classic 4 / Classic FS / Casa Med Ultra / Casa Med Ultra FS / Casa Med Low 01.01.2012
GB – 30