50LBS
2
20KG
2
ld
ld
5LBS
0.002
2KG
0.001
10LBS
0.005
5KG
0.002
20LBS
0.01
10KG
0.005
50LBS
0.02
20KG
0.01
9. After the capacity is set press the
“ZERO”
key. The display will flash
“ ld”
and show the middle
up/down weight to be used in calibrating the scale. This can not be set to 0. Input the amount of
weight to use at the middle up/down calibration.
10. Press the
“ZERO”
key and the display will flash
“u-dP”
and display the weight decimal point. Use
the W-dP from the above chart for the capacity of the scale. Press the
“ZERO”
key.
11. The display will flash
“ld”
and then show the increment. Use the ld from the above chart for the
capacity of the scale. Press the
“ZERO”
key.
12. The display will flash
“dUAL”
and then show a number. 0 disables dual range while 1 enables
dual range. Select 0 or 1. Press the
“ZERO”
key.
13. The scale will flash
“tArE”
and display a number. The
“UNIT”
key changes the number. Only 0
or 1 is allowed. 0 allows for full scale tare. Selecting
“1”
and then pressing the
“ZERO”
key
allows limiting the tare from 0 to full scale. The display will show 5 digits. As an example on a
50LB scale if it is made to be 05000 the scale will only tare up to 4.99.
14. Press the
“ZERO”
key and the display will go back to
“CAL 10”
.
15. If all the calibration settings are done turn the scale off and then on. The display will show the
version number, count up, and go to the weigh mode.