AD-Series

 

 

CALIBRATION MODE WEIGHT SETTINGS FOR SW1N SERIES 

 
 

1.  Make sure the scale is off.  Remove the platter assembly (metal and/or plastic platform.) 
 
2.  On the right hand side you will find a hole and a screw at the bottom.  This is the sealing screw.   

Remove the sealing screw and then turn the scale upside down to find the calibration screw and 
silver plate  located on the right side of the bottom of the scale (if keyboard is facing you and you 
tilt the scale upwards by the front.) 

 
3.  Remove the calibration screw and the silver plate.  Under the plate is the calibration switch. 
 
4.  Press and hold the switch while turning the scale on. The display will show 

“ONE” 

and then 

“CAL

 

1”

 
5. The 

“TARE”

 key advances through the cal settings whereas the 

“UNIT”

 goes down. The 

“ZERO”

 key enters the cal setting. 

 
6. Press the 

“UNIT”

 key twice to go to 

“CAL 10”

. Press the 

“ZERO”

 key to enter into the settings. 

 
7.  The display will flash 

“UNIt”

 and then a number. 0 is for calibrating in KG and 1 is for calibrating in 

LBS. Select the unit that is being used for calibration and press the 

“ZERO”

 key. 

 
8.  The display will flash 

“CAPA”

 and then show the current capacity. The chart below shows the 

capacities to be set. Only use the capacity setting that is shown on the front of the scale. The 

“UNIT”

 key will increment the flashing digit while the 

“TARE”

 key selects the digit that will flash. If    

1 was selected for the unit follow the settings listed under 1. If 0 was selected for the unit follow 
the settings under 0. 

 

 Display 

Shows 

 Display 

Shows 

      Unit 

         1 

 

        0 

  Capacity 

 

  Capacity 

 

   5LBS 

00005 

       2KG 

00002 

 10LBS 

00010 

       5KG 

00005 

 20LBS 

00020 

     10KG 

00010 

 50LBS 

00050 

     20KG 

00020 

                    

ld

 

Weight used 
in Cal 3 
during 
calibration. 
Cannot be 
00000 but 
less then full 
capacity 

This is part 
of a three 
point 
calibration 
during cal 

Weight 
used in Cal 
3 during 
calibration. 
Cannot be 
00000 but 
less then 
full capacity 

 W-dP 

 

W-dP 

 

   5LBS 

       3 

       2KG 

        3 

 10LBS 

       3 

       5KG 

        3 

 20LBS 

       2 

      10KG 

        3 