background image

 

 

30

F31

 

Function

 

 Set COM1 - Output Mode

 

Display Meaning 

F31   0 

 No data out 

F31   1 

 Transmission for both the stable and instable time (stream mode) 

F31   2 

 One time transmission after the weight is stabilized. 

F31   3 

 Transmission only if data is requested. 
 * Data request signal: device ID (F26) _ 1 byte communication 
    (Data on request: 1= 0x01, 10 = 0x0A) 

Setting range 

(0 ~ 4) 

F31   4 

 Response to the data request - Command Mode 

 Set the value of F31 to '1' or more if the print mode is used. 
 
 

Note 1. Command Mode Table

 

 

Data Request Signal of CI-200

 

1  2  3  4

7

10 

11 

Descriptions on 

Request Signal

 

CI-200 

Output Signal

 

dd 

K  Z CR 

LF   

 

 

 

 

 Zero Point Key 

 Received Data Return 

dd 

K  T CR 

LF   

 

 

 

 

 Taro Point Key 

 Received Data Return 

dd 

K  W CR 

LF   

 

 

 

 

 Weight Data Request Signal   Received Data Return 

dd 

I  D

0

0  CR 

LF 

 Device Number 

 Received Data Return 

 

Note 1. (D : 0x44, dd:00~99, K:0x4B , Z:0x5A , CR : 0×0D, LF: 0×0A) 

dd = Device Number (2byte), CR = 0×0D, LF: 0×0A 

          Ex) If a device number is 10, dd becomes 0x31 and 0x30.  
 
Ex) If you want to operate the zero point key when a device number is 11, the indicator operates zeroing if the hex 
code of  “ 44 31 31 4B 5A 0D 0A” is sent. 
 

Содержание PDI Series

Страница 1: ...2 ㅋ ...

Страница 2: ...geroussituations Cautionsaredividedinto Warning and Alert whichmeanasfollows Keepthismanualinaplacewhereproductuserscanfindout afterfinishreadingit Warning means a great possibility led to the death or heavy injury when instructions are violated Alert means a great possibility led to the injury or material damage when instructions are violated ...

Страница 3: ...4 ...

Страница 4: ...5 ...

Страница 5: ...anel Descriptions 13 3 How to Install 15 3 1 How to Connect Load Cell 15 4 Weight Setup Mode 16 4 1 Weight Setup Menu 16 5 Set Mode 23 5 1 How to Enter the Set Mode 23 5 2 Descriptions on key operations in the Set Mode 23 5 3 Set Menu Descriptions F00 F99 24 5 3 1 General Function 26 5 3 2 RS 232 Serial Communication Function 29 5 3 3 Print Function 33 6 Test Mode 34 6 1 How to Enter the Test Mode...

Страница 6: ...RS 232C Port Connection 37 8 2 RS 232 Communication Protocol 37 8 2 1 22 Bytes for CAS 37 8 2 2 10 Bytes for CAS 38 8 2 3 A18 Bytes forAND 38 9 Error Message 39 9 1 ErrorMessagefromtheWeightSetupMode 39 9 2 ErrorMessagefromtheWeighingMode 40 ...

Страница 7: ...tions and various external interfaces which is designed to comply well with special requirements in a variety of industrial fields under strong and beautiful designs in appearance In addition it is designed for the user friendly programs for the easier use of indicator by any user with the built in message display functions to help users understand the product Please use the product right and suff...

Страница 8: ...umweightandadivisionvalueasauserdesires Independentzeroingfunctions Compatible various ECR 1 3 Analog and A D Conversion Applied voltage for load cell DC 5V 350Ω maximum 8 possible connections Zeroing range 0 2mV V 0 5 uV D OIML Ntep KS Input sensitivity 0 5 uV D Non OIML Ntep KS Non straightness 0 01 Full Scale A D internal resolution 1 520 000 1 10 000 NTEP OIML KS A D external resolution 1 20 0...

Страница 9: ...gn for status STABLE NET ZERO COM 1 5 General Specifications AC Adapter AC 100 240 V DC 12V 1 25A Operating temperature 10 40 Product size 141 X 73 X 101 X 382 cf External Dimension Product weight 500g 1 6 Communication and Option Basic COM1 RS 232 Printer PC Interface Optional USB USB is option ...

Страница 10: ...11 2 Specifications in Appearance 2 1 External Dimension PDI ...

Страница 11: ...urrent display of weight is a net weight Zero point The current weight is 0 kg Communication The current status is at the communication status 3 Keyboard It sets the weight display near zero point to 0 Arangeof2 5 10 20 and100 canbeselected Use it to weigh with the tare The current weight is memorized as the tare by pressing the key Press the key when the load tray is empty to release the tare Use...

Страница 12: ...13 2 3 Rear Panel Descriptions ...

Страница 13: ...1 SEALING CAL S W Use it to set the weight calibration POWER Use it for the power supply LOAD CELL A port to connect load cell RS 232C Serial Com 1 connect PC or printer 4 ...

Страница 14: ...tween the load cell output and input sensitivity The input sensitivity of this product is maximum 0 2uV digit or more The following equation should be satisfied upon the system design Applied voltage of load cell x Output voltage of load cell x Value of a division 0 2uV Rated capacity of load cell x Number of load cell Example 1 Number of load cell 4 ea Rated capacity of load cell 500 Kg Rated out...

Страница 15: ...ile pressing Cal S W after removing the sealing Press the setup key long in the weight setup mode to return to the weighing mode 4 1 Weight Setup Menu CAL1 CAL9 CAL 1 Maximum capacity CAL2 Minimum division and decimal position setting CAL 3 Weight calibration 3 1 Setting the range of multiple calibration 3 2 Zero calibration 3 3 Setting weight 3 4 Span calibration CAL7 Gravity adjustment CAL8 Zero...

Страница 16: ...riptions d 0 001 Minimumdivision0 001kg d 0 002 Minimumdivision0 01kg d 0 005 Minimumdivision0 1kg d 0 01 Minimumdivision1kg Save and next Menu navigation Set value change d 0 1 Minimumdivision10kg Note 1 The minimum division means the value of a division Note2 Settheexternalresolutionwithin1 30 000asthevaluebydividingthemaximumweightbytheminimum division Iftheexternalresolutionis1 30 000 or more ...

Страница 17: ...ng places STEP 5 Settingmulticalibrationforstep5 CAL3 3 and CAL 3 4 are carried out five times If the actual curve of load cell is a straight line set the range of weight setup to 1 A function to use when the output of load cell is corrected by setting multiple points in some sections because the actual curve of load cell is not a straight line zero span Zero and Span points to interpolate weight ...

Страница 18: ...e ERR26 isdisplayed CAL3 3 Function SettingWeight Rangeofsetvalue 1 99 999 Used key Display Descriptions LOAD 1 Itmeanstheweightsettingmode Number multicalibrationnumber W 100 00 100 00 unit KgorTon Save and next Menu navigation Set value change Changing places W 0 10 0 10 unit KgorTon Note1 Settheweightwithinarangeof10 100 Although100 ofthemaximumweightisgivenastheinitialvalue enterthedesiredweig...

Страница 19: ...displayed Note4 Whenzeropointishigh anerrormessage ERR25 isdisplayed Note5 Afterfinishingcalibration pressthesetupkeyfor2secondsormoretoconverttotheweighingmode Note6 InordertoenterthenextCALmenu CAL7 10 use ZERO keytoenterthemode CAL7 Function GravityAdjustment Used key Display Descriptions It means you accessed to the menu for the gravityadjustment Gr CAL 9 XXXX Setthegravityfortheproductionplac...

Страница 20: ...message Err27 is displayed Note 4 When zero point is too high an error message Err 26 is displayed CAL9 Function FactorCalibration Used key Display Descriptions NOTUSE This function cannot be used because of multicalibration FACtor It means you entered the factor correction mode Save and next Menu navigation Set value change Changing places 12345 Thecurrentfactorisdisplayed Note 1 As this is a men...

Страница 21: ...ge is displayed and return to the CALmenu mode CAL10 2 Function Setting theappliedsection forthe Dual Range Rangeofsetvalue 0 99999 Used key Display Descriptions M 1000 Dualrangeisappliedtolessthan1000kg M 5000 Dualrangeisappliedtolessthan5 000kg Save and next Menu navigation Set value change Changing places M 10000 Dualrangeisappliedtolessthan10 000kg Note 1 If the input value is greater than the...

Страница 22: ... key at the indicator front to start the Set Mode After finishing the setup in the Set Mode press key for a long time turn off the power and turn on the power again 5 2 Descriptions on key operations in the Set Mode Save and Next Menu navigation Set value change Changing places ...

Страница 23: ...rationFilter F07 02 MotionDetectionCondition F08 02 AutomaticZeroTrackingCompensation F09 00 WeightBackup F13 10 SetZeroRange F14 01 SetZERO TARE KeysAvailability F17 01 Set Key F21 10 SetInitialZeroRange F24 00 SetBacklightOperationalCondition LCD F25 03 SetBacklightBrightness Note Number in is the default at the factory shipment ...

Страница 24: ...F30 00 COM1OutputFormat F31 00 COM1 OutputMode F32 04 COM2BaudRate F33 01 COM2Usage F34 00 COM2 OutputFormat F35 00 COM2 OutputMode F36 01 Set ECRType PrintFunction F43 01 PrintLineFeed SetModeInitialization F99 SettheSetvalueofSetModetotheFactoryDefault Note Number in is the default at the factory shipment ...

Страница 25: ...e 00 50 F05 50 DisplayofaverageforNo 50 F06 Function Settingvibrationfilter Display Meaning F06 00 VibrationfilterOFF F06 10 Compensationforthevibrationvalueof5divisions 0 5d 10 Settingrange 00 99 F06 99 Compensationforthevibrationvalueof49 5divisions 0 5d 99 Note 1 Apply this function to a place with heavy vibrations The display response speed becomes slower when the vibration filter is applied N...

Страница 26: ... WeightBackupFunction Display Meaning F09 0 Weightbackupisnotused Settingrange 0 1 F09 1 Weightbackupisused Note 1 As the backup state memorizes the initial status at zero for the weighing machine even during the blackout or when the power is turned off the weight value is displayed if there is any weighing object in the weighing machine when the power is turned on If the weighing tray is empty pr...

Страница 27: ...ightoff F24 1 Backlightonwhenanykeyisoperated F24 2 Backlightonwhentheweightchanges F24 3 Backlightonwhenitis Stable aftertheweightchanges F24 4 Backlightonwhenakeyoperatesortheweightchanges Settingrange 0 5 F24 5 Backlightonallthetime Note Although it is set to 5 press the power key shortly to turn off the backlight F25 Function SetBacklightandLEDBrightness Display Meaning F25 1 Set10 ofbrightnes...

Страница 28: ... stopbit2 paritybit oddnumber Note1 F26andF27applycommonlyto2serialcommunications RS23CandPRT Serial Communication COM1 Function F28 Function SetCOM1BaudRate Display Meaning F28 0 600bps F28 1 1200bps F28 2 2400bps F28 3 4800bps F28 4 9600bps F28 5 19200bps F28 6 38400bps F28 7 57600bps Settingrange 0 8 F28 8 115200bps F29 Function SetCOM1 Usage Display Meaning F29 0 Connecttoaprinter Settingrange...

Страница 29: ...used Note 1 Command Mode Table Data Request Signal of CI 200 0 1 2 3 4 5 6 7 8 9 10 11 Descriptions on Request Signal CI 200 Output Signal D dd K Z CR LF ZeroPointKey ReceivedDataReturn D dd K T CR LF TaroPointKey ReceivedDataReturn D dd K W CR LF WeightDataRequestSignal ReceivedDataReturn D dd I D 0 0 0 0 0 CR LF DeviceNumber ReceivedDataReturn Note1 D 0x44 dd 00 99 K 0x4B Z 0x5A CR 0 0D LF 0 0A ...

Страница 30: ...playedwithnoprint COM1andCOM2cannotbeusedtogetherastheprinterfunction F34 Function SetCOM2 OutputFormat Display Meaning F34 0 22bytesforCA F34 1 10bytesforCA Settingrange 0 2 F34 2 18bytesforAND F35 Function SetCOM2 OutputMode Display Meaning F35 0 Nodataout F35 1 Transmissionforboththestableandinstabletime streammode F35 2 Onetimetransmissionaftertheweightisstabilized F35 3 Transmissiononlyifdata...

Страница 31: ...6 6 SAMSUNGECR SPAIN B 9600 D 8 S 1 P NONE F36 7 AP1 Type1_ECRsendtoscale11H B 9600 D 8 S 1 P NONE F36 8 ELZABDelta B 9600 D 8 S 1 P EVEN F36 9 ELZABALFA B 9600 D 8 S 1 P EVEN F36 10 MEDESAS100 ECR B 9600 D 8 S 1 P EVEN F36 11 MEDESAS100 ECR answerforENQ B 9600 D 8 S 1 P EVEN F36 12 MEDESAS100 ECR continuoustransmision B 9600 D 8 S 1 P EVEN F36 13 MEDESAS100 ECR answerforENQ B 9600 D 8 S 1 P EVEN ...

Страница 32: ... kg Net 0 999 kg No 0003 Gross 0 999 kg Tare 0 000 kg Net 0 999 kg No 0004 Gross 0 999 kg Tare 0 000 kg Net 0 999 kg Note 1 If the power is turned off and then on the number and total are initialized to 0001 Note 2 The possible number for print is a range of 1 9999 F99 Function Setdefault Display Meaning 0 Noinitializationfunctionsforindicator Settingrange 0 1 1 Carryouttheinitializationfunctionsf...

Страница 33: ...ST mode press key for a long time turn off the power and turn on the power again 6 2 Test Menu TEST 1 TEST4 Test 1 Key test Test 2 Display test Test 3 Load cell test andA D conversion test Test 4 RS 232 serial communication test COM1 COM2 Test 1 Function Key test Used key Display Descriptions Test value change Test Move to next menu KEY 1 When you press any key to test the number and code for the ...

Страница 34: ...s fixed or 0 is displayed check the connection of load cell once again Test 4 Function Serial Communication Test Used key Display Descriptions Test value change Test Move to next menu Tx Rx 05 13 Status to wait for transmission or reception Transmission 5 Reception 13 Note 1 Run this test while the communication program in the computer ex Hyper Terminal is executing after connecting a serial port ...

Страница 35: ...ght including the tare cannot exceed the maximum weight Put a containeron the load tray Containerweight 10kg Press thetarekey Tareissaved Put an object on the load tray Net weight 20kg Press the total net weight key thevalueof object s weight is displayed Removethecontainerand objectfromtheload trayto displaythesaved tare If the tare is removed Removethecontainerand objectfromthe load tray andpres...

Страница 36: ...rt female RS 232C port of CI 200 Serial port of the computer 8 2 RS 232 Communication Protocol 8 2 1 22 Bytes for CAS 1 Data bit 8 Stop bit 1 Parity bit none 2 Code ASCII 3 Set the time to send data to the computer in the Set Mode Send all the time if F30 and F35 are set to 1 Send when the weight is stable if F30 and F35 are set to 2 Send upon data request if F30 and F35 are set to 3 Only if the c...

Страница 37: ...mal forexample 13 5kg 8bytesofASCIIcode correspondingto0 0 0 0 1 3 and 5 aresent Lamp Status Byte Bt7 1 Bt6 Stable Bt5 0 Bt4 Hold Bt3 Printer Bt2 Gross Weight Bt1 Tare Bt0 ZeroPoint 8 2 2 10 Bytes for CAS 1 Databit 8 Stopbit 1 Paritybit none 2 Code ASCII 3 Transmissiondataformat 10bytes Data 8bytes CR LF 8 2 3 18 Bytes for AND 1 Databit 7 Stopbit 1 Paritybit oddnumber evennumber 2 Code ASCII 3 Tra...

Страница 38: ...morethan100 ofthe maximumcapacity Settheweightwithinthemaximumcapacity setinCAL1 from CAL4intheweightsetupmode Err24 Toolowspan Settheweightagainbyloweringtheresolutionasthesettingofthe currentresolutionisnotpossiblebecauseofeitherabnormalityor loweroutputintheloadcell TwolowweightforPCSandpercentsample Err25 Toohighspan Thereiseitheranyabnormalityortoohighoutputintheloadcell Executestepsfromtheze...

Страница 39: ...atF14intheSetMode Err09 Thecurrentweightisoutoftherangeofzero point Settherangeofoperationsforthezerokeytowithin 2 or10 atF13intheSetMode Err10 Thetaretosetisoutofthemaximumweightof thescale Setthetaretolessthanthemaximumweight Err13 Thesetvalueofzeropointontheweightsetting isoutofrange Checkthestatusoftheloadtrayandsettheweight again Err15 Therangeexceededduringsettingtheitemcode inthecommandmode...

Страница 40: ...IICodeTable Character Code Character Code Character Code Character Code Character Code Character Code Space 32 0 48 64 P 80 96 p 112 33 1 49 A 65 Q 81 a 97 q 113 34 2 50 B 66 R 82 b 98 r 114 35 3 51 C 67 S 83 c 99 s 115 36 4 52 D 68 T 84 d 100 t 116 37 5 53 E 69 U 85 e 101 u 117 38 6 54 F 70 V 86 f 102 v 118 39 7 55 G 71 W 87 g 103 w 119 40 8 56 H 72 X 88 h 104 x 120 41 9 57 I 73 Y 89 i 105 y 121 ...

Страница 41: ...42 MEMO ...

Страница 42: ...43 MEMO ...

Страница 43: ...44 MEMO ...

Страница 44: ...45 ...

Отзывы: