ER
PLUS
Service Manual
15
9/27/2010
(6) Press [TARE] key to set the value, and then the scale displays “CAPA“ and “15”. It means
that full-capability is “15 (calibration unit, kg or lb)”
(7) Input a capability by using [TARE] key.
(8) And then “mid“ message is displayed. “5” will be shown. It means that mid-capability is “5
(calibration unit, kg or lb)”
(9) Input a mid-capability by using keypad.
(10) Press [TARE] key, and then “W-dP “message and “3” will be shown. It means that weight
decimal point is “3 (will display 0.000)”
(11) Input a weight decimal point by keypad.
(12) Press [TARE] key, and then “1d “ message and “0.005” will be shown. It means that the
one division is “0.005 (calibration unit, kg or lb)”
(13) Input an one division by using keypad.
(14) Press [TARE] key, and then “dual” message and “1” will be shown. It means that dual
interval is enabled. (0 : disable, 1 : enable)”
(15) Input a dual interval setting by using keypad.
(16) Press [TARE] key, and then “tare” message and “0” will be shown. You can
enable or
disable custom tare (0:disable, 1:enable)
(17) Press [TARE] key to save calibration factor and then “C-10 ” message will be shown. Add
to this you can browse any menu to reset value by using [ZERO]key and [-] key in the C-10
Mode.
* CAL-10 TABLE
CAPA(Kg)
Interval
Resolution
CAL 10 SETTING
Unit
CAPA
Mid W-dp 1d dual
tare
1
Dual
1/2000
3
1000
500
1
0.5
1
0 or 1
2
Single
1/10000
3
2000
1000
1
0.2
0
0 or 1
2
Dual
1/2000
0
2
1
3 0.001 1
0 or 1
3
Dual
1/3000
0
3
1
3 0.001 1
0 or 1
5
Single
1/10000
3
5000
2000
1
0.5
0
0 or 1
5
Dual
1/2500
0
5
2
3 0.002 1
0 or 1
6
Dual
1/3000
0
6
3
3 0.002 1
0 or 1
10
Single
1/10000
0
10
5
3 0.001 0
0 or 1
10
Dual
1/2000
0
10
5
3 0.005 1
0 or 1
12
Dual
1/2400
0
12
5
3 0.005 1
0 or 1
15
Dual
1/3000
0
15
5
3 0.005 1
0 or 1
20
Single
1/10000
0
20
10
3 0.002 0
0 or 1
20
Dual
1/2000
0
20
10
2 0.01
1
0 or 1
25
Dual
1/2500
0
25
10
2 0.01
1
0 or 1
Содержание ER PLUS LCD
Страница 1: ...ERPLUS Service Manual 1 9 27 2010 ERPLUS SERVICE MANUAL ...
Страница 8: ...ERPLUS Service Manual 8 9 27 2010 1 9 Dimension 107 222 384 413 ...
Страница 9: ...ERPLUS Service Manual 9 9 27 2010 1 10 Sealing Method ...
Страница 10: ...ERPLUS Service Manual 10 9 27 2010 ...
Страница 20: ...ERPLUS Service Manual 20 9 27 2010 3 The Schematics and Diagram 3 1 System Block Diagram ...
Страница 21: ...ERPLUS Service Manual 21 9 27 2010 4 Exploded View CAS CO LAB 2003 12 29 FOR APPROVAL ...
Страница 22: ...ERPLUS Service Manual 22 9 27 2010 CAS CO LAB 2004 01 29 FOR APPROVAL ...
Страница 23: ...ERPLUS Service Manual 23 9 27 2010 5 Load Cell drawing 25 112 15 112 SW spring element 2090501 00 ...
Страница 28: ...ERPLUS Service Manual 28 9 27 2010 z DRAWING NUMBER 9000 ERS 0010 ERPLUS LCD REAR DISPLAY ...
Страница 33: ...ERPLUS Service Manual 33 9 27 2010 z DRAWING NUMBER 9000 ER1 0010 ERPLUS VFD POLE DISPLAY TYPE 2 ...
Страница 38: ...ERPLUS Service Manual 38 9 27 2010 6 1 4 2 Front Display PCB Part Location z DRAWING NUMBER 6110 PER 1001 0 ...
Страница 39: ...ERPLUS Service Manual 39 9 27 2010 6 1 4 3 Rear Display PCB Part Location z DRAWING NUMBER 6110 PER 1002 0 ...
Страница 41: ...ERPLUS Service Manual 41 9 27 2010 6 1 5 2 Front Display PCB Part Location z DRAWING NUMBER 6110 PER 1100 0 ...
Страница 42: ...ERPLUS Service Manual 42 9 27 2010 6 1 5 3 Rear Display PCB Part Location z DRAWING NUMBER 6110 PER 1101 0 ...
Страница 44: ...ERPLUS Service Manual 44 9 27 2010 6 1 6 2 KEY PCB Part Location Pb z DRAWING NUMBER 9000 ERS 0006 KEY PCB ...