background image

 

 

 

 

INST-PM-CRSY102 

Rev B   04/11/2018 

  

 

2.

 

Retrieve old antimicrobial copper rod from inside reservoir, squeeze spring clip 
and slide off antimicrobial copper rod as shown below. 

 

 

3.

 

Discard old antimicrobial copper rod and spring clip.  

4.

 

Slightly squeeze new spring clip together. Then insert long white end of new 
antimicrobial copper rod into new spring clip as shown below. 

 

Antimicrobial Copper Rod  

Spring Clip 

Detached Antimicrobial Copper Rod  

Antimicrobial Copper Rod  

Содержание CRSY102

Страница 1: ...CRSY102 INST PM CRSY102 ...

Страница 2: ...Revision Log Revisions Description Letter Date A 01 03 2017 Initial release B 04 11 2018 New cover ...

Страница 3: ...ntenance Kit for Model CRSY102 are shown below 1 Turn off recirculator and unplug power cord 2 If necessary pull recirculator away from wall INSTRUCTIONS FOR REPLACING CARBON FILTER DI CARTRIDGE ANTIMICROBIAL COPPER ROD AND ATOMIZER IN MODEL CRSY102 REPLACING DI CARTRIDGE Disconnect electrical power before removing access panel s Carbon Filter DI Cartridge Silicone Grease Antimicrobial Copper Rod ...

Страница 4: ...access panels Remove panels by pulling down on latch tabs rotating panel slightly outward and lifting as shown below 4 Remove fastening screw and slide filter wrench off DI cartridge housing as shown below Access Panel Latch Tabs Fastening Screw Filter Wrench ...

Страница 5: ...RSY102 Rev B 04 11 2018 5 Unscrew DI cartridge housing on left side of recirculator using filter wrench as shown below Note Drain water from DI cartridge housing into bucket Fastening Screw DI Cartridge Housing ...

Страница 6: ...INST PM CRSY102 Rev B 04 11 2018 6 Remove and discard old DI cartridge Filter Wrench Do not discard O ring inside of DI cartridge housing ...

Страница 7: ...ntire 0 ring with silicone grease as shown below Note Do not use more than half of silicone grease Ensure tapered end of DI cartridge is located next to O ring groove Ensure O ring is seated inside groove in DI cartridge housing O ring Groove Tapered End of DI Cartridge Silicone Grease ...

Страница 8: ...turn further as shown below 1 Unscrew carbon filter housing on right side of recirculator using filter wrench as shown below Note Drain water from carbon filter housing into bucket 2 Remove and discard old carbon filter REPLACING CARBON FILTER Filter Wrench Carbon Filter Housing Do not discard O ring inside of carbon filter housing ...

Страница 9: ...rculator until hand tight and use filter wrench to tighten 1 8 turn further as shown below 5 Slide filter wrench onto DI cartridge housing and fasten screw 1 Unscrew reservoir cap located on top of recirculator as shown below REPLACING ANTIMICROBIAL COPPER ROD Ensure O ring is seated inside groove in filter housing Carbon Filter Housing Reservoir Cap ...

Страница 10: ... antimicrobial copper rod as shown below 3 Discard old antimicrobial copper rod and spring clip 4 Slightly squeeze new spring clip together Then insert long white end of new antimicrobial copper rod into new spring clip as shown below Antimicrobial Copper Rod Spring Clip Detached Antimicrobial Copper Rod Antimicrobial Copper Rod ...

Страница 11: ... below 6 Squeeze spring clip while inserting antimicrobial copper rod as shown below 7 Place new antimicrobial copper rod inside reservoir and replace cap 1 Locate lower reservoir on left side of recirculator as shown below REPLACING ATOMIZER Lower Reservoir Antimicrobial Copper Rod ...

Страница 12: ...INST PM CRSY102 Rev B 04 11 2018 2 Unsnap 3 lid tabs on lower reservoir and lift to locate atomizer as shown below Lid Tabs Atomizer ...

Страница 13: ... counterclockwise using 5 8 wrenches as shown below 4 Install new atomizer by turning nozzle clockwise using 5 8 wrenches 5 Reinstall lid and snap 3 lid tabs in place as shown below Note Do not snap tab located under filter wrench as shown below Atomizer Filter Wrench Snapped Lid Tabs ...

Страница 14: ...hes catch 7 If necessary roll recirculator back into desired location and reconnect chamber drain s to recirculator inlet 8 Plug in power cord and turn on recirculator Chamber drain recirculator inlet is gravity fed Any vertical rise or kink in drain line that is higher than chamber drain will result in a trap and not allow chamber to drain to recirculator ...

Отзывы: