INST-PM-7545-2-UVA
Rev A 04/16/2018
4.
Remove humidity diffuser by lifting out of chamber as shown below.
5.
Remove and discard old antimicrobial copper rod by squeezing spring clip and
sliding antimicrobial copper rod out of clip as shown below.
6.
Slightly squeeze spring clip and insert long white end of new antimicrobial copper
rod into clip.
Note: Clip must be oriented as shown below after installing rod.
7.
Reinstall humidity diffuser with antimicrobial copper rod towards front of
chamber.
8.
Reinstall access panel.
9.
Secure access panel with retained (6) screws using Phillips screwdriver bit.
10.
Reinstall bottom shelf.
11.
Continue to instructions for replacing additional accessory preventative
maintenance items if applicable or plug in power cord and turn on chamber.
Humidity Diffuser
Antimicrobial Copper Rod
Spring Clip
Spring Clip Bend Flush with Humidity Diffuser