background image

 

 

 

 

INST-PM-6045 

Rev B   04/04/2018 

  

 

7.

 

Remove (2) screws securing drip pan using Phillips screwdriver bit as shown 
below.  

Note:  Retain screws.

 

 

 
 
 

8.

 

Pull drip pan downward as shown below.  

Note:  If necessary use flatblade 

screwdriver to pry drip pan down as shown below.  

 

  

9.

 

Remove and discard old atomizers using (2) 8″crescent wrenches as shown below.

 

 

 
 

 

10.

 

Install new atomizers. 

11.

 

Reinstall drip pan.  

Note:  If necessary use flatblade screwdriver to install drip pan.  

 

12.

 

Secure drip pan with retained (2) screws using Phillips screwdriver bit. 

13.

 

Align tabs on intake diffuser with drip pan. 

14.

 

Secure intake diffuser with retained (4) screws using Phillips screwdriver bit. 

15.

 

Reinstall false floor. 

Drip Pan Mounting Screws 

Drip Pan 

Drip Pan 

Atomizer 

Atomizers 

Содержание 6045

Страница 1: ...MODEL 6045 INST PM 6045 ...

Страница 2: ...Revision Log Revisions Description Letter Date A 09 05 2017 Initial release B 04 04 2018 New cover ...

Страница 3: ... may be necessary Parts included in Preventative Maintenance Kit for Model 6045 are shown below Note Photos may have a different appearance depending on chamber options 1 Remove air filter cover by pulling outward on both sides near bottom as shown below INSTRUCTIONS FOR REPLACING AIR FILTER AND ATOMIZERS IN MODEL 6045 REPLACING AIR FILTER Air Filter Cover Air Filter Atomizers ...

Страница 4: ...ed edge of air filter cover as shown below Note Orient cardboard side of air filter towards chamber 4 Install air filter cover with angled edge at top while aligning air filter with guide tracks on both sides and snap into place as shown below Note Right side is shown in photos Air Filter Guide Track Angled Edge on Air Filter Cover Air Filter Seam ...

Страница 5: ... from chamber by pulling access port upwards while bending edges downward as shown below 5 Remove 4 screws securing intake diffuser using Phillips screwdriver bit as shown below Note Retain screws 6 Remove intake diffuser by pulling towards front of chamber REPLACING ATOMIZERS Disconnect electrical power before removing interior parts Intake Diffuser Mounting Screws as viewed from inside chamber u...

Страница 6: ...d discard old atomizers using 2 8 crescent wrenches as shown below 10 Install new atomizers 11 Reinstall drip pan Note If necessary use flatblade screwdriver to install drip pan 12 Secure drip pan with retained 2 screws using Phillips screwdriver bit 13 Align tabs on intake diffuser with drip pan 14 Secure intake diffuser with retained 4 screws using Phillips screwdriver bit 15 Reinstall false flo...

Страница 7: ...Rev B 04 04 2018 16 Reinstall shelf channels 17 Reinstall shelves 18 Continue to instructions for replacing additional accessory preventative maintenance items if applicable or plug in power cord and turn on chamber ...

Отзывы: