Symbol Descriptions 

 

The CE mark and Notified Body 
Registration Numbers, the requirement of 
Annex II from Medical Device Directive 
93/42/EEC are met.

 

 

Indicates this device is subject to the Waste 
Electrical and Electronic 
Equipment Directive in the European Union. 
To protect the environment, dispose of 
useless device at appropriate collection sites 
according to national or local regulations. 

 

Authorized 
representative in 
the European 
community 

 

Caution 

 

Please read the instructions for use 

 

Manufacturer 

 

BF type applied part 

 

Battery Recycling 

 

Paper Recycling 

IP22

 

Classification for water ingress and 

particulate matter. 

 

 

 

 

 

Guidance and manufacturer’s declaration – electromagnetic emissions

 

The MFT-9 is intended for use in the electromagnetic environment specified below. The customer or the user of the MFT-9 should assure that it is 
used in such an environment.

 

Emissions test

 

Compliance

 

Electromagnetic environment 

– 

guidance

 

RF emissions 
CISPR 11

 

Group 1

 

The MFT-9 uses RF energy only for its internal function. Therefore, its RF emissions are 
very low and are not likely to cause any interference in nearby electronic equipment. 

 

RF emissions 
CISPR 11

 

Class B

 

The MFT-9 is suitable for use in all establishments, including domestic establishments 
and those directly connected to the public low-voltage power supply network that supplies 
buildings used for domestic purposes.

 

Harmonic emissions 
IEC 61000-3-2

 

Not applicable

 

Voltage fluctuations/ 
flicker emissions 
IEC 61000-3-3

 

Not applicable

 

 

Guidance and manufacturer’s declaration – electromagnetic immunity 

The MFT-9 is intended for use in the electromagnetic environment specified below. The customer or the user of the MFT-9 should assure that it is 
used in such an environment. 

Immunity test 

IEC 60601 test level

 

Compliance 

level

 

Electromagnetic environment 

– 

guidance

 

Conducted RF 
IEC 61000-4-6 
 
 
Radiated RF 
IEC 61000-4-3 

3 Vrms 
150 kHz to 80 MHz 
 

 

3 V/m 
80 MHz to 2,5 GHz 

Not 

applicable 

 
 
 

3  V/m 

Portable and mobile RF communications equipment should be used no closer to any 
part of the MFT-9, including cables, than the recommended separation distance 
calculated from the equation applicable to the frequency of the transmitter. 

 
Recommended separation distance

                                                                                                                                                                                           

d = 

1,2 

                                                                                                                                                                           

d = 

1,2 

  80 MHz to 800 MHz

                                                                                                                                                                                   

d = 

2,3 

  800 MHz to 2,5 GHz 

 

where 

is the maximum output power rating of the transmitter in watts (W) according 

to the transmitter manufacturer and 

is the recommended separation distance in 

meters (m). 
 
Field strengths from fixed RF transmitters, as determined by an electromagnetic site 
survey, a should be less than the compliance level in each frequency range b. 

 

Interference may occur in the vicinity of equipment marked with the following symbol: 

 

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. 
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption 
and reflection from structures, objects and people. 

Содержание MFT-9

Страница 1: ...re If in doubt about the person s condition seek medical assistance YELLOW Caution Prompt response by the OPERATOR is required GREEN Ready for use Memory Locations There are total 9 sets memories for body temperature Each memory also records the measurement date time mode icon and site location When in power on press the ON MEM button to see the temperature stored with icon F C Switch In power off...

Страница 2: ...you hear a beep sound Remove the probe from ear and read the temperature measurement on the LCD After each ear measurement wait the ear icon stop flashing to be ready for next measurement 4 2 Forehead temperature measuring points for attention Remarks a Forehead temperature is displayed in oral mode This mode converts the forehead temperature to display its oral equivalent Value b Before the measu...

Страница 3: ...sage Problem Solution Error 5 9 the system is not functioning properly Unload the battery wait for 1 minute and repower it If the message reappears contact the retailer for service Measurement before device stabilization Wait until all the icons stops flashing The ambient temperature is not within the range between 50o F and 104o F 10o C 40o C Allow the thermometer to rest in a room for at least 1...

Страница 4: ... domestic purposes Harmonic emissions IEC 61000 3 2 Not applicable Voltage fluctuations flicker emissions IEC 61000 3 3 Not applicable Guidance and manufacturer s declaration electromagnetic immunity The MFT 9 is intended for use in the electromagnetic environment specified below The customer or the user of the MFT 9 should assure that it is used in such an environment Immunity test IEC 60601 test...

Страница 5: ... for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 sec Not applicable Mains power quality should be that of a typical commercial or hospital environment If the user of the MFT 9 requires continued operation during power mains interruptions it is recommended that the MFT 9 be powered from an uninterruptible power supply or a battery Power frequency 50 60 Hz magnetic field IEC 61...

Отзывы: