Symbol Descriptions
The CE mark and Notified Body
Registration Numbers, the requirement of
Annex II from Medical Device Directive
93/42/EEC are met.
Indicates this device is subject to the Waste
Electrical and Electronic
Equipment Directive in the European Union.
To protect the environment, dispose of
useless device at appropriate collection sites
according to national or local regulations.
Authorized
representative in
the European
community
Caution
Please read the instructions for use
Manufacturer
BF type applied part
Battery Recycling
Paper Recycling
IP22
Classification for water ingress and
particulate matter.
Guidance and manufacturer’s declaration – electromagnetic emissions
The MFT-9 is intended for use in the electromagnetic environment specified below. The customer or the user of the MFT-9 should assure that it is
used in such an environment.
Emissions test
Compliance
Electromagnetic environment
–
guidance
RF emissions
CISPR 11
Group 1
The MFT-9 uses RF energy only for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The MFT-9 is suitable for use in all establishments, including domestic establishments
and those directly connected to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Not applicable
Guidance and manufacturer’s declaration – electromagnetic immunity
The MFT-9 is intended for use in the electromagnetic environment specified below. The customer or the user of the MFT-9 should assure that it is
used in such an environment.
Immunity test
IEC 60601 test level
Compliance
level
Electromagnetic environment
–
guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2,5 GHz
Not
applicable
3 V/m
Portable and mobile RF communications equipment should be used no closer to any
part of the MFT-9, including cables, than the recommended separation distance
calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
d =
1,2
d =
1,2
80 MHz to 800 MHz
d =
2,3
800 MHz to 2,5 GHz
where
P
is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer and
d
is the recommended separation distance in
meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site
survey, a should be less than the compliance level in each frequency range b.
Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.