AF0118 | 2018-03-27
8-1
8
Appendix A
Optional Parts
Dose Area Product (DAP)
DAP is available as an option DRX-Revolution Nano Mobile X-ray System. The DAP is
configured and calibrated via the monitor display for the use of a DAP measurement
chamber and additional filters. The X-ray tube will support DAP with automatic
population. DAP meter specifications are the following:
Measurement
Specification
Upper limit of Response Range
0.8 nC/cGy-cm2
Saturation
98% at 400 V, 1 Gy/s
Energy Range
40 – 150 kV
Energy Dependence
5% over the 40–110 kV range
Chamber Filter Effect
0.20 mm A1 (70 kV)
0.25 mm A1 (70 kV)
Wireless infrared remote exposure switch
The X-ray tube will support an optional infrared wireless exposure switch. See the
User
Guide for Remote Exposure Switches for Mobile X-ray Systems
for instructions on using
this tool.
RFID Badge Reader
See
DirectView Online Help Topics.