30 Chapter 5 Acquiring Images using the Cephalometric Modality
Preparing and Positioning the Adult and Pediatric Patient
To prepare and position the patient, follow these steps:
1.
Ask the patient to remove all metal objects.
2.
Ask the patient to wear a lead apron. Ensure that the apron lays flat across the patient’s shoulders.
3.
Pull out at the
top part
of a head clamp bar to extend both bars.
4.
Ask the patient to do the following:
•
Move forward.
•
Stand up straight in the following positions:
•
For a frontal AP, facing the generator.
•
For a frontal PA, facing the cephalometric sensor.
•
Place both ears in between the head clamps.
5.
On the control panel, press and hold
to level the ear cones to the patient’s auditory canals.
Important: The head clamp will NOT extend if you pull out at the lower
part of the bar.