
38
The calibration error for expiration and
inspiration at each flow rate are
displayed. The calibration error should
be less that 3%. If a greater error is
shown, repeat the procedure ensuring
that the syringe is emptied and filled in
a smooth manner without jerking the
handle. If an error greater than 3% is
still
shown,
inspect
the
turbine
transducer and clean if necessary.
Содержание MicroLoop
Страница 2: ......
Страница 7: ...5 XI 220 223 224 229 231 232 233 SPCS 233 235 236 237 238 Customer Contact Information 241...
Страница 8: ......
Страница 30: ......