Chapter 5 Operation
29
675–101–101 Rev. P
2. Alarm set/confirm Screen
Figure 10 Alarm set/confirm Screen
Touch the NCPAP button or the Alarm Mute / Reset button for 3 seconds to set the
alarms and to move to the next screen. If either button is not touched within 2
minutes, the alarm limits will be set automatically. When the alarm limits have been
set, the screen display changes to the Mode Select Screen with the driver operating
in NCPAP mode.
Note:
Press High Flow meter must be checked through the Start-up screen and be set
during operation. This enables manual breath or back-up apnea breath, where
applicable, to be active.
CAUTION
Do not adjust any of the Alarm Limit settings to an extreme value. Selecting an
extreme value can make the alarm system inadequate for the patient.
Figure 11 Mode Select Screen
Содержание Infant Flow SiPAP
Страница 1: ...Infant Flow SiPAP Operator s Manual ...
Страница 10: ...x Infant Flow SiPAP 675 101 101 Rev P Electromagnetic Environment Specifications ...
Страница 11: ...Infant Flow SiPAP xi 675 101 101 Rev P ...
Страница 14: ...xiv Infant Flow SiPAP 675 101 101 Rev P Infant Flow SiPAP ...
Страница 20: ...6 Chapter 2 Product Specifications 675 101 101 Rev P Infant Flow SiPAP ...
Страница 38: ...24 Chapter 5 Operation 675 101 101 Rev P Rear Panel Figure 8 Rear Panel ...
Страница 48: ...34 Chapter 5 Operation 675 101 101 Rev P Infant Flow SiPAP ...
Страница 64: ...50 Appendix A Product Configurations 675 101 101 Rev P Infant Flow SiPAP ...
Страница 65: ...Infant Flow SiPAP 51 675 101 101 Rev P Appendix B Pneumatic Diagram ...
Страница 66: ...52 Appendix B Pneumatic Diagram 675 101 101 Rev P Infant Flow SiPAP ...
Страница 70: ...56 Appendix C Alarm Troubleshooting 675 101 101 Rev P Infant Flow SiPAP ...
Страница 76: ...62 Appendix D Fault Management 675 101 101 Rev P Infant Flow SiPAP ...
Страница 78: ...64 675 101 101 Rev P Infant Flow SiPAP ...