TRACh Team
19
PREPARING FOR TRANSPORT
Steps for Assembling the Circuit for transport
1.
Wash your hands with soap and water for 15 seconds and dry your hands with a clean towel.
2.
Gather your supplies.
3.
Turn heater off.
4.
Remove the heater wires from the temperature probe ports on the circuit.
5.
Remove the short tubing from the bacterial filter that connects to the ventilator.
6.
Remove the angled end of the long tubing from the water chamber and connect to the bacterial filter.
7.
Briefly check your child for chest rise to make sure the ventilator is functioning correctly.
The LTV or/and ReVel ventilator images(s) is/are © 2016 CareFusion Corporation; Used with permission.
Be sure to attach a HME in line with
the ventilator circuit for transport
.
Attach angled end of ventilator circuit to the bacterial filter.
Attach the 3 clear tubes to the right side of the ventilator.