Bird Sentry™ Blender
Chapter 7: Appendices
L1383 Rev. D
21
Note:
All specifications assume the following standard environmental conditions unless specified otherwise.
●
Ambient and sample gas temperatures of 77°F (25°C)
●
Barometric pressure of 30Hg (760mm Hg)
●
Ambient relative humidity of 50%
●
Sample gas relative humidity of 0%
Disposable Galvanic O
2
Sensors
Expected useful life
750,000 O
2
% hours* (approx. two-years average use)
Shelf life
The date by which the sensor must be placed into service to achieve
one year service life is marked on each sensor package.
Sensor warm-up time
One minute, nominal time required after the initial installation of
sensor before specifications are applicable.
Setup and Installation
Battery Installation
Remove the battery cover located at the left side facing out by inserting a screwdriver (at location
noted) and prying. Observing correct polarity, install two (2) standard "AA" 1.5V alkaline batteries and
replace cover.