CareFusion AVEA Скачать руководство пользователя страница 5

Mechanics (continued)

P

tp

 PEEP

The difference between the corresponding airway and esophageal 
pressures at the end of the expiratory hold during  
an AutoPEEP maneuver. Requires an inspiratory hold and 
esophageal catheter.

MIP

The maximum negative airway pressure that is achieved by the 
patient, during an expiratory hold maneuver.

P

100

The negative pressure that occurs 100 ms after an inspiratory 
effort has been detected.

WOB

V

Ventilator work of breathing (WOB

V

) is defined as the summation 

of airway pressure (PAW) minus the baseline airway pressure 
(PEEPAW) times the change in tidal volume to the patient (-V) 
during inspiration, and normalized to the total inspiratory tidal 
volume (Vti).

WOB

P

Patient work of breathing (WOB

P

), normalized to the total 

inspiratory tidal volume. Patient work of breathing is defined as 
the summation of two work components: Work of the lung and 
work of the chest wall. Requires an esophageal balloon.

WOB

I

Imposed work of breathing (WOBI) is defined as the work 
performed by the patient to breathe spontaneously through the 
breathing apparatus, (i.e., the E.T. tube), the breathing circuit and 
the demand flow system. Requires a tracheal catheter. 

Note:

 Monitored values are displayed as BTPS.

Capnography

EtCO

2

The patient’s peak expired CO

2

 as measured and reported by the 

CO

2

 sensor in the airway, calculated for each breath then averaged 

as specified by set EtCO

2

 averaging time.

VCO

2

Minute volume of exhaled CO

2

 measured continuously and 

averaged over a user-selectable time. Requires flow measurement 
at the wye or circuit compliance active.

VtCO

2

Tidal volume of exhaled CO

2

, measured for each breath and then 

averaged over the set VCO

2

 averaging time. Requires flow mea-

surement at the wye or circuit compliance active.

Vd ana

The volume of dead space in the patient’s conducting airways 
from the nose to the level of the terminal bronchioles measured 
for each breath, then averaged over the set CO

2

 averaging time. 

Also includes any mechanical dead spaces added to the ventila-
tor circuit between the CO

2

 sensor and the patient. Requires flow 

measurement at the wye or circuit compliance active.

Vd/Vt ana

Anatomic Vd/Vt is averaged over the set VCO

2

 averaging 

time. Requires flow measurement at the wye or circuit  
compliance active.

Vd phy

Comprises Vd ana as well as the volume of the respiratory zone 
(respiratory bronchioles, alveolar ducts and alveoli) not participat-
ing in gas exchange. Requires an arterial blood gas sample.

Vd/Vt phy

Physiologic Vd/Vt is averaged over the set VCO

2

 averaging time. 

Requires an arterial blood gas sample.

Vd alv

Alveolar dead space is the difference between physiological  
dead space and anatomical dead space. It represents the volume 
of the respiratory zone that is from ventilation of relative  
under-perfused or non-perfused alveoli. Requires an arterial 
blood gas sample.

Monitors and definitions (continued)

Содержание AVEA

Страница 1: ...in screen B Event Select events to record to trends C Alarm silence 2 minute silence period D Suction 3 functions 1 O2 maneuver for 2 minutes 2 Disables demand flow on loss of PEEP 3 Alarms silenced for 2 minutes E Nebulizer 20 minutes synchronized with inspiration B C E D A User interface ...

Страница 2: ... in addition to the internal inspiratory flow sensor and heated expiratory flow sensor The variable orifice sensor attaches to the receptacle circled in dark blue The hot wire flow sensor attaches to the receptacle circled in light blue directly below the variable orifice flow sensor connection To attach pull back the locking collar Push firmly onto the receptacle Push the collar forward to lock t...

Страница 3: ...ed by the patient during the last minute Ve kg Minute volume adjusted for patient weight Spon Ve Spontaneous minute volume Spon Ve kg Spontaneous minute volume adjusted for patient weight Rate time monitors Rate Breath rate Spon rate Spontaneous breath rate Ti Inspiratory time Te Exhalation time I E Inspiratory expiratory ratio Note Not active for demand breaths f Vt Rapid shallow breathing index ...

Страница 4: ...atient circuit and the tracheal sensor Requires an inspiratory hold and tracheal catheter RLUNG The ratio of the tracheal pressure differential peak plateau to the inspiratory flow 12 ms prior to the end of inspiration Requires an inspiratory hold and tracheal catheter dPAW The difference between peak airway pressure PPEAK AW and base line airway pressure PEEPAW dPES The difference between peak es...

Страница 5: ... breath then averaged as specified by set EtCO2 averaging time VCO2 Minute volume of exhaled CO2 measured continuously and averaged over a user selectable time Requires flow measurement at the wye or circuit compliance active VtCO2 Tidal volume of exhaled CO2 measured for each breath and then averaged over the set VCO2 averaging time Requires flow mea surement at the wye or circuit compliance acti...

Страница 6: ... OI Oxygenation index is a dimensionless number often used to assess the pressure cost of oxygenation calculated from the FIO2 mean airway pressure and an arterial blood oxygen measurement required entered by the clinician Monitors and definitions continued Available modes Mode Adult Pediatric Neonatal Volume A C X X X Volume SIMV X X X Pressure A C X X X Pressure SIMV X X X TCPL A C X TCPL SIMV X...

Страница 7: ... 05 to 0 12 0 15 Airway resistance 2 to 5 cmH2O L s 15 cmH2O L s Lung compliance 50 to 100 mL cmH2O 25 mL cmH2O Maximum inspiratory 30 cmH2O low effort 20 cmH2O Pressure MIP 140 cmH2O high effort Auto PEEP 0 3 cmH2O Respiratory drive P0 1 2 to 4 cmH2O 6 cmH2O f VT 60 to 90 105 Pressure time 200 to 300 cmH2O sec min Product PTP Ti TTOT 0 3 to 0 4 Parameter Acceptable range SaO2 90 on FlO2 of 40 P A...

Страница 8: ...nel in evaluating weaning potential If measured values exceed acceptable range successful weaning may be less likely Ranges from these parameters are not intended as a substitute for clinical assessment by qualified medical personnel and CareFusion assumes no liability for their use in patient care A list of references is available upon request CareFusion Germany 234 GmbH Leibnizstrasse 7 97204 Ho...

Отзывы: