5B-26
CORRECTIVE MAINTENANCE
PCA Module
Alaris
®
Syringe Module (8110 Series)
Alaris
®
PCA Module (8120 Series)
Technical Service Manual
5.2.22 Removing Linear Sensor Assembly
1. Peel Linear Sensor Assembly off
Extrusion Frame.
2. Use alcohol to clean adhesive debris
from Extrusion Frame.
During Reassembly:
1. Ensure Extrusion Frame surface is
clean.
2. Before adhering Linear Sensor
Assembly to Extrusion Frame:
a. Remove protective backing from
Linear Sensor.
b. Align hole in Linear Sensor
Assembly with hole in Extrusion
Frame.
3. Use a burnishing tool to ensure that
Linear Sensor Assembly is securely
adhered to Extrusion Frame.
4. Clean Linear Sensor Assembly surface
using alcohol and lint-free cloth (such as
Kimwipes or lint-free tissue).
5.2
DISASSEMBLY/REASSEMBLY
(Continued)
To avoid damaging the Linear Sensor
Assembly, do not touch with bare hands.
CAUTION
Alignment Hole
Extrusion Frame
Linear Sensor Assembly
Содержание Alaris 8110 series
Страница 9: ...1 GENERAL INFORMATION...
Страница 10: ......
Страница 15: ...2 CHECKOUT CONFIGURATION...
Страница 16: ......
Страница 27: ...3 PREVENTIVE MAINTENANCE...
Страница 28: ......
Страница 31: ...4 PRINCIPLES OF OPERATION...
Страница 32: ......
Страница 41: ...5A CORRECTIVE MAINTENANCE SYRINGE MODULE...
Страница 42: ......
Страница 71: ...5B CORRECTIVE MAINTENANCE PCA MODULE...
Страница 72: ......
Страница 105: ...6 TROUBLESHOOTING...
Страница 106: ......
Страница 115: ...7A ILLUSTRATED PARTS BREAKDOWN SYRINGE MODULE...
Страница 116: ......
Страница 143: ...7B ILLUSTRATED PARTS BREAKDOWN PCA MODULE...
Страница 144: ......
Страница 171: ......