5A-13
CORRECTIVE MAINTENANCE
Syringe Module
Alaris
®
Syringe Module (8110 Series)
Alaris
®
PCA Module (8120 Series)
Technical Service Manual
5.2.11 Removing Pressure Sensor Board
Assembly
1. Disconnect Pressure Sensor Harness
from Pressure Sensor Board.
2. Remove four (
4
) screws attaching Base
Disk Holder to Top Disk Holder and
Front Case.
3. Remove Base Disk Holder and Pressure
Sensor Board Assembly.
During Reassembly:
When installing an old version of Base Disk
Holder, use
4-40 x
5
/
16
screws (item
486
) in
upper mountings, and
4-40 x
3
/
16
screws
(item
484
) in lower mountings. When
installing a newer version of Base Disk
Holder, use
4-40 x
5
/
16
(item
486
) for all
5.2
DISASSEMBLY/REASSEMBLY
(Continued)
Pressure Sensor Board Assembly
Pressure Sensor
Harness
Base Disk Holder (old)
To avoid damage to the Front Case when
reinstalling the Base Disk Holder, use
correct size screws (reference Figure
7A-4
).
CAUTION
Base Disk Holder (new)
(See CAUTION statement)
No boss
Boss
4-40 x
5
/
16
(4 PL)
4-40 x
5
/
16
(upper mountings)
4-40 x
3
/
16
(lower mountings)
Содержание Alaris 8110 series
Страница 9: ...1 GENERAL INFORMATION...
Страница 10: ......
Страница 15: ...2 CHECKOUT CONFIGURATION...
Страница 16: ......
Страница 27: ...3 PREVENTIVE MAINTENANCE...
Страница 28: ......
Страница 31: ...4 PRINCIPLES OF OPERATION...
Страница 32: ......
Страница 41: ...5A CORRECTIVE MAINTENANCE SYRINGE MODULE...
Страница 42: ......
Страница 71: ...5B CORRECTIVE MAINTENANCE PCA MODULE...
Страница 72: ......
Страница 105: ...6 TROUBLESHOOTING...
Страница 106: ......
Страница 115: ...7A ILLUSTRATED PARTS BREAKDOWN SYRINGE MODULE...
Страница 116: ......
Страница 143: ...7B ILLUSTRATED PARTS BREAKDOWN PCA MODULE...
Страница 144: ......
Страница 171: ......