5A-9
CORRECTIVE MAINTENANCE
Syringe Module
Alaris
®
Syringe Module (8110 Series)
Alaris
®
PCA Module (8120 Series)
Technical Service Manual
5.2.7
Removing Logic Board Assembly
NOTE:
If the Logic Board must be replaced
on a Syringe Module with version
7
or earlier
software, the Logic Board must be replaced
at the depot. See the note at the beginning
of this chapter for more information.
1. If present, remove Retainer Clip from
Dual Connector Cable-Logic Board
connection.
2. Disconnect Dual Connector Cable from
Logic Board
J8
.
3. Disconnect Linear Sensor harness from
Logic Board
J12
.
4. Remove one (
1
) screw attaching Keypad
ground wire and Logic Board to
IUI
Support Bracket.
5. Remove one (
1
) screw attaching Logic
Board to Internal Frame Assembly.
6. Disconnect harnesses from following
locations on Logic Board:
J4
(display)
J7
(motor)
J11
(flange detect)
J14
(pressure sensor)
7. Remove Logic Board.
During Reassembly:
Route Dual Connector Cable and Linear
Sensor Harness under
IUI
Support Bracket.
5.2
DISASSEMBLY/REASSEMBLY
(Continued)
Dual Connector Cable
Logic Board Assembly
Internal Frame Assembly
Keypad Ground Wire
Retainer Clip
Linear Sensor Harness
Содержание Alaris 8110 series
Страница 9: ...1 GENERAL INFORMATION...
Страница 10: ......
Страница 15: ...2 CHECKOUT CONFIGURATION...
Страница 16: ......
Страница 27: ...3 PREVENTIVE MAINTENANCE...
Страница 28: ......
Страница 31: ...4 PRINCIPLES OF OPERATION...
Страница 32: ......
Страница 41: ...5A CORRECTIVE MAINTENANCE SYRINGE MODULE...
Страница 42: ......
Страница 71: ...5B CORRECTIVE MAINTENANCE PCA MODULE...
Страница 72: ......
Страница 105: ...6 TROUBLESHOOTING...
Страница 106: ......
Страница 115: ...7A ILLUSTRATED PARTS BREAKDOWN SYRINGE MODULE...
Страница 116: ......
Страница 143: ...7B ILLUSTRATED PARTS BREAKDOWN PCA MODULE...
Страница 144: ......
Страница 171: ......