Maximum output of optical radiation for the “SAD Light”:

The “SAD Light ” is manufactured by:

Zero-Plus International Limited

Room 1004, 10/F., Join-In Hang Sing Centre,

71-75 Container Port Road, Kwai Chung,

New Territories, Hong Kong

Email: [email protected]

The “SAD Light” is distributed by:

European Authorised Representative:

Globalmind Consumer Electronics GmbH

Ernst-Mantius-Str.11, 

21029 Hamburg, 

Germany

 Important

If this appliance is damaged in any way or it is not working and you cannot find the answer to your question in FAQ. Please 

contact customer service or the retailer for assistance. 

Radiance Output

Risk group classified by IEC 

60601-2-57:2011

Max. value

E

UVA

: Eye UV-A

Exempt Group

1.0x10

-4  

W

.

m

-2

ES: Actinic UV skin & eye

Exempt Group

1.0x10

-4  

W

.

m

-2

EIR: Infrared radiation hazard 

exposure limits for the eye 

Exempt Group

1.0x10

1  

W

.

m

-2

EH: Thermal hazard for the skin

Exempt Group

2.0x10

1  

W

.

m

-2

LB: Blue light

Exempt Group

5.0x10

1  

W

.

m

-2.

sr

-1

LR: Retinal thermal

Exempt Group

5.0x10

1  

W

.

m

-2.

sr

-1

12. Contact information

Содержание DL08127

Страница 1: ...Usage and Maintenance Instructions www careco co uk CareCo Hubert Road Brentwood Essex CM14 4JE SAD Day Light DL08127 Rev 201905 ...

Страница 2: ... and winter whilst eliminating the danger ous UV rays contained in sunlight The additional light helps adjust the levels of melatonin that can help the symptoms of SAD Seasonal Affective Disorder Seasonal Affective Disorder SAD as known as winter depression or winter blues It may cause people Feel low listless and fatigued Need for more sleep Have trouble concentrating Less active than normal Lack...

Страница 3: ...r supply lines 1kV signal input output 100 kHz repetition frequency 2kV for power supply lines Not applicable 100 kHz repetition frequency Surge IEC 61000 4 5 0 5kV 1kV differential lines 0 5kV 1kV 1kV common mode 0 5kV 1kV differential lines Not Applicable Voltage dips interruptions and variations on power supply input lines IEC 61000 4 11 0 UT 0 5 cycle At 0 45 90 135 180 225 270 and 315 0 UT 1 ...

Страница 4: ... TETRA 400 Pulse modulation 18 Hz 1 8 0 3 27 450 430 470 GMRS 460 FRS 460 FM 5 kHz deviation 1 kHz sine 2 0 3 28 710 704 787 LTE Band 13 17 Pulse modulation 217 Hz 0 2 0 3 9 745 780 810 800 960 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulation 18 Hz 2 0 3 28 870 930 1720 1700 1990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS Pulse modulation 217 Hz 2 0 3 28 1845 1970 2...

Страница 5: ...e Important Safety note indicating possible dam age to the appliance accessory Caution Consult accompanying docu ments Specifies serial number of the SAD Light Class II Equipment OFF ON Date of Manufacture Authorized Representative in the Euro pean Community Manufacturer Ingress Protection Rating The following symbols appear in this manual on the SAD Light or on its accessories 3 Signs and symbols...

Страница 6: ... is plugged in do not allow any water to be sprayed onto it The appliance must be operated only when it is completely dry Protect the appliance from strong impacts Avoid pulling the cord when removing the plug from the socket Instead hold the plug If the adaptor cord or the appliance is damaged contact customer service or the dealer since special tools are needed to make repairs Never try to repai...

Страница 7: ...munications equipment including peripherals such as antenna cables and external antennas should be used no closer than 30 cm to any part of this appliance including power cable Otherwise degradation of the performance of this equipment could result If the appliance does not light on or flickering due to the electromagnetic disturbances do not use it and try to place the appliance in other location...

Страница 8: ... the hole on the back of appliance 3 Place the appliance on a solid surface approximately 20 30cm from your eyes the light must bathe your face from the side 4 Connect the appliance only to the mains voltage listed on the type plate Note Push the plug completely into the mains socket Ensure that there is a socket near the place where the appliance will stand Lay the mains cable in such a way that ...

Страница 9: ...hese should pass If in doubt just lower your treatment times until comfortable again Do not leave the appliance unattended in the duration IV Enjoying light over longer periods During the darkest period of the year repeat the treatment for at least 7 consecutive days or longer depending on your individual requirements V Switching off Press the ON OFF button of the appliance to switch off Warning B...

Страница 10: ...f separately If you have any questions please contact your local authorities responsible for waste disposal 9 Disposal 10 FAQ Problem Possible causes Solutions Device does not light up On Off button switched off Switch on the on off button No Power connect the main plug correctly No Power The main cable is defective Contact customer services or your retailer LEDs have reached the end of its servic...

Страница 11: ...umidity Atmospheric pressure 700 1060hPa Storage and transport environment 10 to 60 C and 30 to 90 relative humidity Atmospheric pressure 700 1060hPa Product classification 1 Class II equipment 2 No applied part 3 Protection against ingress of water IP21 4 Not category AP APG equipment 5 Mode of operation Continuous operation Maximum length of power cord 2000 mm Location of emission aperture Scree...

Страница 12: ...ius Str 11 21029 Hamburg Germany Important If this appliance is damaged in any way or it is not working and you cannot find the answer to your question in FAQ Please contact customer service or the retailer for assistance Radiance Output Risk group classified by IEC 60601 2 57 2011 Max value EUVA Eye UV A Exempt Group 1 0x10 4 W m 2 ES Actinic UV skin eye Exempt Group 1 0x10 4 W m 2 EIR Infrared r...

Отзывы: