4.8.
Pressure gradient
According to the methodology of lymphatic drainage mas-
sage, the pressure value in each subsequent chamber
should be lower than the value in the preceding chamber.
CarePump series of devices has
„gradient”
function
A
,
which allows the user to precisely set the pressure drop in
the subsequent chambers. To activate this function, select
it from the main screen. Then the gradient settings menu
will be displayed
B
.
The initial pressure
is the pressure value in the first air-
-inflating chamber.
The pressure drop
is the value by which the pressure in
the chambers will drop gradually, starting from the value
set in “starting pressure”, up to the value specified in “fi-
nal pressure”. The pressure drop can be set in the pressure
unit: mmHg or as a percentage value. It can be selected by
switching between
“mmHg”
and the symbol
“%”
in the
parameter “pressure drop”.
Final pressure
is the pressure value that will be obtained in
the last chamber filled with air.
The selected gradient settings can be confirmed by press-
ing
“OK”
button. To turn the gradient off, press
“OFF”
but-
ton or set the value of “pressure drop” to 0.
GRADIENT
OK
OFF
Initial pressure
20
mmHg
Pressure drop
Final pressure
1 mmHg
17
mmHg
1
Lymphoedema I
mmHg
min
level
20
65
5
3
3
Time
Pressure
Speed
Interval Hold
Gradient
1
2
3
4
20
18
17
GRADIENT
OK
OFF
Initial pressure
20
mmHg
Pressure drop
Final pressure
1 mmHg
17
mmHg
1
Lymphoedema I
mmHg
min
level
20
65
5
3
3
Time
Pressure
Speed
Interval Hold
Gradient
1
2
3
4
20
18
17
A
B