Reuse and cleaning
The product is reusable. Follow the instructions
below for cleaning and reconditioning. It is
recommended to clean the mattress between
users. Consult your hygiene manager or Care of
Sweden for help and instructions if you are
unsure. Cleaning and reconditioning should
otherwise be performed as required.
8.1
Cleaning
CONTROL UNIT
Wipe off the control unit with a damp
cloth and mild cleaning solution for this
purpose: such as alcohol with or without
tenside surfactants, oxidizing solutions, chlorine
max 1 % or hydrogen peroxide max 1,5 %.
If using a different agent, choose one that does
not affect the control unit.
MATTRESS COVER
Covers can be wiped off with a mild
cleaning solution for this purpose such
as: alcohol with or without tenside
surfactants, Isopropanol 70 %, oxidizing
solutions, chlorine max 1 % (max 10% for cover
Stone) or hydrogen peroxide max 1,5 %.
Excrement and blood stains must be removed
as soon as possible using cold water. Carefully
follow local instructions and the instructions for
the detergent.
Multi-layer covers should be divided before
washing. The parts should be washed with
similar colours.
For chlorine, we recommend a mixture
of max 1% (max. 10% for cover Stone).
INNER COVER
Clean the affected area with a mild cleaning
solution for this purpose such as: alcohol with
or without tenside surfactants, Isopropanol 70
%, oxidizing solutions, chlorine max 1 % or
hydrogen peroxide max 1,5 %.
MATTRESS
1. Disconnect the tube connector from the
control unit and remove the air from the
mattress.
2. Remove the covers.
3. Move the mattress to a clean area that is
suitable for cleaning.
4. Wipe off the cells, all tubes and the CPR
module with a mild cleaning solution for this
purpose such as: alcohol with or without
tenside surfactants, Isopropanol 70 %, oxidizing
solutions, chlorine max 1 % or hydrogen
peroxide max 1,5 %. Let it dry.
5. Wipe off the work surfaces using a suitable
detergent and/ or disinfectant.
6. Put the mattress together. If cells for any
reason have become loose from the tubes,
these must be replaced according to the
drawing in section 7.2.
FOAM CORE
Clean the affected area with a mild detergent
(such as washing-up liquid) and water or with
an alcohol-based disinfectant (cleaner intended
for this purpose. Gently squeeze out any water).
Note!
•
When using the CuroCell® CX15 or
CuroCell® CX20, the air cells can be divided
using a quick-release connector, making it
easier to clean the cells.
Note!
•
Check the hygiene cover, cells and hoses
each time the product is cleaned. If
damaged, it must be replaced or repaired.
Also check the control unit, tube connectors
and power cable during cleaning. Damaged
parts must be replaced or repaired.
•
Do not wring or roll the foam core to
extract the water. Let it dry in a warm,
ventilated area (not in direct sunlight). The
foam core must be completely dry before it
is used again.