Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The xr50+ ergometer is intended for use in the electromagnetic environment specifi ed below. It is the responsibility of the
customer or user to ensure that the xr50+ ergometer is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment –
Guidance
Electrostatic discharge (ESD) to
EN 61000-4-2
± 8 kV contact
± 16 kV air
± 8 kV
± 16 kV
Floors should be wood, concrete or
ceramic tile. If fl oors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Electrical fast transient / burst
to EN 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input and
output lines
± 2 kV passed
Mains power should be that of a typical
commercial or hospital environment.
Surge to EN 61000-4-5
± 1 kV diff erential mode
± 2 kV common mode
± 1 kV
N. A.
Mains power should be that of a typical
commercial or hospital environment.
Voltage dips, short interrup-
tions and voltage variations
on power supply input lines to
EN 61000-4-11
to EN 61000-4-11
passed
Mains power should be that of a typical
commercial or hospital environment. If
the user of the xr50+ ergometer requires
continued operation during power mains
interruptions, it is recommended that
the xr50+ ergometer be powered from
an uninterruptible power supply or a
battery.
Power frequency (50/60 Hz)
magnetic fi eld to EN 61000-4-8
30 A/m
50 Hz
passed
Power frequency magnetic fi elds should
be at levels characteristic of a typical
location in a typical commercial or hos-
pital environment.
The xr50+ ergometer has no compo-
nents susceptible to magnetic fi elds.
Note:
UT is the a.c. mains voltage prior to application of the test level.