DISPLAY CODES
CALIBRATION REQUIRED -
Indicates improper stored calibration data,
calibration is necessary.
OVER CAPACITY -
Indicates a weight exceeding the capacity of scale has
been placed on the scale tray.
ERROR -
NOV-RAM test failure.
ANALOG FAILURE -
This could mean a damaged or defective load cell, a
failed component in the analog circuity of the scale or the platform isn’t on or is
on incorrectly.
SETUP AND CALIBRATION
Your scale has been thoroughly tested and calibrated before being shipped to you and should
not require adjustment. In the event the scale should need re-calibration, the following
calibration procedure should be performed. To maintain the scale’s high degree of accuracy, it
is recommended that a qualified technician with certified test weights re-calibrate the scale.
In order to perform Setup and Calibration, it is necessary to remove the "Do Not Remove"
sticker from the lower right side of the scale. With the sticker removed, locate the Setup and
Calibration switch in the opening closest to the front of the scale. Refer to the figure on the
next page.
BEGIN SETUP and CALIBRATION
1. Making certain the scale tray is empty, press and hold the setup and calibration switch
located on the lower right side of the scale. Refer to the figure on the next page.
2. Press the
ON/TARE
key to turn the scale on. Release the setup and calibration switch
when the display is showing
6&$
.
SCALE MODEL
The display will show
6&$
. Press the hidden
key to display the current setting. For a PS-6
the display should show
3
. Press the hidden
key until
3
is displayed. Press the
ON/TARE
key to save the setting.
CALIBRATION
The display will show
&$/
. Press the hidden
key to display the current setting. To calibrate
the scale, press the hidden
key until
<(6
is displayed, then press the
ON/TARE
key. The
display will change to
/2G
.
To skip calibration, press the hidden
key until
QR
is displayed, then press the
ON/TARE
key to continue the setup.
LOAD WEIGHT
With
the display will showing
/2G
, place a (2) lb certified test weight on the scale tray.
Allow approximately five seconds for the scale to stabilize. Press the
ON/TARE
key to start
the calibration. Dashes will be displayed from the left to the right of the display
momentarily, then the display will change to
X/G
.
3
NOTE: If operating from batteries, install the batteries making certain that
they are positioned in accordance with the drawing on the circuit board. If
using the optional AC power adapter, insert the plug into the jack on the
back of the scale.
2&
(UU
&$/
(UD