2-64
Programming
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
Multidose Mode
(Continued)
CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
ALARM
INFUSE
STANDBY
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
1.
If
Current time
displayed is correct, press
CONFIRM
soft
key; otherwise, press
Change Time
and enter correct
time. (Reference "System Options", "Time of Day" in PC
Unit Section of this
DFU
2.
Press
VOLUME DURATION
soft key.
3.
To enter volume to be infused for each dose, use numeric
data entry keys.
4.
To enter duration for each dose, press
DURATION
soft key
and use numeric data entry keys.
5.
To enter time interval (
1
to
24
hours) between doses,
press
DOSE INTERVAL
soft key and use numeric data
entry keys.
6.
To enter number of doses, press
# OF DOSES
soft key
and use numeric data entry keys.
• If Delay Options is enabled,
DELAY OPTIONS
soft key
7.
To begin multidose infusion, press
START
soft key.
• Main Display shows remaining
VTBI
for that dose.
• At completion of a multidose program,
MULTIDOSE
COMPLETE
appears on Main Display.
Multidose
A
Time of Day
Change
Time
Current time:
08:00
CONFIRM
>CONFIRM Time-of-Day
Volume/Duration Enabled
Содержание Alaris 8000 Series
Страница 8: ......
Страница 52: ......
Страница 54: ......
Страница 176: ......
Страница 242: ......
Страница 284: ......
Страница 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Страница 318: ......
Страница 338: ......
Страница 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Страница 340: ......
Страница 350: ......
Страница 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Страница 352: ......
Страница 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Страница 362: ......
Страница 377: ......
Страница 378: ......