Table 4: Guidance and Manufacturer's Declaration—Electromagnetic Immunity
The Air
Life
®
nCPAP System Driver is intended for use in the electromagnetic environment
specified below. The customer or user of the Air
Life
®
nCPAP System Driver should assure
that it is used in such an environment.
Immunity Test
IEC 60601-1-2
Test Level
Compliance
Level
Electromagnetic Environment—
Guidance
Portable and mobile RF
communications equipment should
be used no closer to any part of
the Air
Life
®
nCPAP System
Driver, including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended Separation
Distance
IEC 61000-4-6
Conducted RF
3 V rms
150 kHz to 80 MHz
3 V rms
d
1.2
P
=
IEC 61000-4-3
Radiated RF
3 V/m
80 MHz to 2.5 GHz
3 V/m
d
1.2
P
=
80 MHz to 800 MHz
E
1
d
2.3
P
=
800 MHz to 2.5 GHz
Where
P
is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and
d
is
the recommended separation
distance in meters (m).
a
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
should be less than the
compliance level in each
frequency range.