Powerheart® AED G3 Pro 9300P
70-00969-01 F
3-7
Step 4: Analyze the heart rhythm
During the analysis phase, you may hear one or more of these prompts:
When the AED prompts…
Do this…
“Do not touch patient. Analyzing rhythm.”
The AED begins analyzing the cardiac
rhythm of the patient.
1.
Do not touch the patient.
2.
Wait for the next prompt.
If the AED prompts…
This is the problem…
Do this…
“Press pads firmly to
patient’s bare skin.”
The pads are not properly
placed or are loose.
Ensure that pads are
firmly placed on clean,
dry skin.
“Analysis interrupted.
Stop patient motion.”
The AED restarts the
analysis.
The patient is excessively
jostled or there is a strong
electromagnetic emitting
device nearby (within 5
meters).
Remove the electronic
device or stop the
excessive motion.