Pads
Version 112-2025-109 B
3-6
1
With the label on the battery facing the AED battery compartment, insert
the battery as shown in the drawing.
2
Push the latched end of the battery firmly into the AED, as shown in the
drawing, until the battery snaps into place. The exposed side of the
battery should be flush with the outside of the AED case.
3
For the Powerheart AED, open the lid for 5 seconds to initiate self-test. If
the battery is installed properly, the STATUS INDICATOR will turn
GREEN. Close the lid.
OR
Open the lid for 5 seconds to initiate a self-test. If the battery is installed
properly, the SMARTGAUGE battery indicator LEDs will illuminate;
additionally, the STATUS INDICATOR will turn GREEN. If service is
required, then the SERVICE indicator will illuminate.
Pads
The defibrillation pads come in a ready-to-use, sealed package containing
one pair of self-adhesive pads with an attached cable and connector. The
pads are disposable and should be discarded after each rescue.