Powerheart® AED G3 9300A and 9300E
70-02006-01 A
2-5
Intellisense battery
Caution: Use only Cardiac Science Approved Equipment.
Using batteries, pads, cables, or optional equipment other than those
approved by Cardiac Science may cause the AED to function improp-
erly during a rescue.
Battery operating life
The battery operating life depends on the type of battery, actual usage
and environmental factors.
A new battery provides:
◆
A minimum of 14 hours (18 hours typical) of device operating
time at 20°C to 30°C ambient temperature with no shocks
delivered, OR
◆
A minimum of 9 hours of device operating time at 0°C
ambient temperature with no shocks delivered, OR
◆
A minimum of 9 hours of device operating time at 50°C ambient
temperature with no shocks delivered, OR
◆
up to 290 shocks (typical)
!
Table 2-1: Normal battery life
Full Operational
Estimated Shelf
Replacement Guarantee
Life (from date of (from date of
Model
manufacture)
installation)
Typical Shocks
9146 Lithium
5 Years
4 years
up to 290
Note:
The battery operating life depends on the type of battery,
device settings, actual usage, and environmental factors.