HTR 

 

 
MF30 – 3.07 

2.0

 

I

NSTALLATION

 

2.1

 

Unpacking & Handling 

 

When unpacking or moving the furnace lift it by both ends of its base. Never lift it by any protruding parts 
or by the lid. Preferably, use two people to carry the furnace. 

 

Remove any packing material from around or inside the furnace before use. 

NOTE: This product contains 

Refractory Ceramic Fibre 

(better described as A

lumino Silicate Wool

for precautions and advice in handling this material see the ‘Repairs and Replacements’ section. 

2.2

 

Siting 

 

Place the furnace in a well ventilated room, away from other sources of heat, and on a surface which is 
resistant to accidental spillage of hot materials. Do not mount the furnace on an inflammable surface. 

 

Ensure that there is free space around the furnace. Do not obstruct any of the vents in the control section: 
they are needed to keep the controls cool. 

 

Ensure that the furnace is placed in such a way that it can be quickly switched off or disconnected from the 
electrical supply - see below. 

2.3

 

Electrical Connections 

 

Connection by a qualified electrician is recommended 

 

The furnace is made for single phase A.C. supply, which may be Live to Neutral non-reversible, 
Live to Neutral reversible or Live to Live. 

 

Check the furnace rating label before connection. The supply voltage should agree with the voltage 
on the label, and the supply capacity should be sufficient for the amperage on the label. 

 

The supply should be fused at the next size equal to or higher than the amperage on the label. A 
table of the most common fuse ratings is also given in section 8.1 of this manual. Where a supply 
cable is present there are internal supply fuses; customer fusing is preferred but not essential. 

 

Furnace with supply cable: either wire directly to an isolator or fit with a line plug. 

 

Furnace without supply cable: a permanent connection to a fused and isolated supply should be 
made to the internal terminals after temporary removal of the furnace back panel. 

 

Connection by line plug: the plug should be within reach of the operator, and should be quickly 
removable. 

 

Connection to isolating switch: this should operate on both conductors (single phase) or on all live 
conductors (three phase), and should be within reach of the operator. 

 

The supply MUST incorporate an earth (ground). 

CONNECTION DETAILS 

supply type

           

 

Supply 

Terminal label 

Cable colour 

Live-Neutral

 

Reversible or Live-Live

 

1-phase

 

    L 

Brown 

To live 

to either power conductor 

 

    N 

Blue 

To neutral 

to the other power conductor 

 

    PE 

Green/Yellow 

To earth (ground) 

to earth (ground) 

 

2.4

 

Gas Supply 

 

The furnace is supplied with a Nitrogen inlet and flowmeter as standard. The gas supply must be 
provided with an on/off valve and pressure regulator to a maximum of 4 psi. Connect the gas 
supply to the inlet on the left-hand side. The exhaust box on the right-hand side may be connected 
to a outlet pipe provided this does not restrict the gas flow. The exhaust incorporates a safety 
pressure relief valve set to relieve pressure above 1 psi in case the exhaust becomes blocked. 

 

Do not use toxic gases without taking appropriate precautions. The furnace is not suitable for use 
with combustible gases. 

Содержание HTR 11/150

Страница 1: ...onnected to the electricity supply Manuals are supplied separately for the furnace controller and overtemperature controller when fitted Please read the controller manuals before operating the furnace...

Страница 2: ...e control circuit is energised Heat Light the adjacent light glows or flashes to indicate that power is being supplied to the elements DANGER of electrical shock read any warning printed by this symbo...

Страница 3: ...n the amperage on the label A table of the most common fuse ratings is also given in section 8 1 of this manual Where a supply cable is present there are internal supply fuses customer fusing is prefe...

Страница 4: ...and tighten part 1 with the spanner Do not overtighten The end of the vessel with this assembly fitted becomes the left hand end Lightly lubricate the drive coupling assembly and the furnace drive wh...

Страница 5: ...temperature controller according to the instructions supplied To turn the furnace off set the Instrument switch to it s off position the controller display will go blank If the furnace is to be left o...

Страница 6: ...ature This is a recrystallisation process which is greatly affected by surface conditions and therefore by contaminants in the furnace atmosphere At 1000 C only a highly contaminated atmosphere has a...

Страница 7: ...all or fax often enables a fault to be diagnosed and the necessary spare part despatched Each furnace has its own record card at Carbolite In all correspondence please quote the serial number model ty...

Страница 8: ...urther information on request Alternatively our service division can quote for any repairs to be carried out at your premises or ours 5 3 Temperature Controller Replacement 301 This controller is fitt...

Страница 9: ...no vessel in place Make a careful note of the colours and positions of all the electrical connections to the inner cylindrical chamber and thermocouple Disconnect the thermocouple and element connecti...

Страница 10: ...or faulty controller There are no lights glowing on the controller Check the supply fuses and any fuses in the furnace control compartment The controller may be faulty or not receiving a supply due to...

Страница 11: ...Reactor 208V 240V Instrument Switch coil temperature controller thermocouple overtemp controller thermocouple SSR e l e m e n t s heat on F1 F2 N L PE Filter if fitted contactor or relay overtemperat...

Страница 12: ...trument Switch coil temperature controller thermocouple overtemp controller thermocouple SSR e l e m e n t s heat on F1 F2 N L PE Filter if fitted contactor or relay overtemperature if fitted motor co...

Страница 13: ...HTR MF30 3 07 13...

Страница 14: ...n board 20mm x 5mm Other 32mm x 6mm F3 Heat Light fuses May be omitted up to 25 Amp phase supply rating 2 Amps glass type F 32mm x 6mm Customer fuses Required if no supply cable fitted Recommended if...

Страница 15: ...e the setting beyond the value in the table the wiring and the elements are designed on the basis of the declared settings Model Volts 200V 208V 220V 230V 240V 110V 115V 120V HTR 11 75 100 100 89 81 7...

Страница 16: ...n of all Furnace and Oven products please contact Carbolite Engineering Services Telephone 44 0 1433 624242 Fax 44 0 1433 624243 Email service carbolite com MF30 3 07 14 08 14 9 0 SPECIFICATIONS Carbo...

Отзывы: