Installation, Operation and Maintenance Instructions

1200°C E-Range Tube Furnace - EVT Model: 150mm
No Controller

EVT 12/150B + No Controller

MEN-EVT12150B-005_NOCTRL

(

30-01-2018

)

Содержание EVT 12/150B

Страница 1: ...Installation Operation and Maintenance Instructions 1200 C E Range Tube Furnace EVT Model 150mm No Controller EVT 12 150B No Controller MEN EVT12150B 005_NOCTRL 30 01 2018...

Страница 2: ...stand 9 2 3 3 Figure Mounting the furnace onto the vertical stand 9 2 3 4 Figure Tube furnace positioned horizontally on the stand 10 2 3 5 Figure Tube support 11 2 3 6 Figure Tube support assembled 1...

Страница 3: ...6 1 Safety Warning Disconnection from Power Supply 21 6 2 Safety Warning Refractory Fibre Insulation 21 6 3 Temperature Controller Replacement 21 6 4 Solid state Relay Replacement 22 6 5 Thermocouple...

Страница 4: ...to the elements Heat switch the switch disconnects power to the heating elements unless this switch is OFF there is a danger of electric shock when inserting objects into the product 1 2 General Warn...

Страница 5: ...ork tube adaptors or it could be possible for an operator to access electrically live element coils that could cause serious injury or death 8 Warning label Switch off the furnace before loading and u...

Страница 6: ...ements are ener gized 4 Product rating label indicating the product serial number and electrical information 5 Access holes to front panel fixing 6 Warning label Disconnect the mains supply before rem...

Страница 7: ...quipment is mounted should be stable and not subject to movement or vibrations The height of the mounting surface is important to avoid operator strain when loading and unloading samples Ensure that t...

Страница 8: ...stand upright to the base using the 6 x M6 dome cap screws 4 and wash ers 3 supplied l Partially screw in the top M6 screw knobs 2 x 6 into the frame upright as shown in 2 3 3 l Hook the furnace onto...

Страница 9: ...bling the stand Index to numbered items in Figure 2 3 2 and Figure 2 3 3 1 Vertical stand base 2 Vertical stand upright 3 Fixing screws M6 x 16 dome cap screw 4 Washer M6 5 Tube furnace 6 M6 screw kno...

Страница 10: ...t the work tube with the furnace in the horizontal position as shown in figure 2 3 4 To fit the work tube in a vertical orientation follow the steps below Omit the use of the end seal if not required...

Страница 11: ...using item 6 until the tube is secure enough for use in a vertical arrangement Fitting arrangement is shown in figure 2 2 l Figure 2 2 shows an exploded view of all the required parts l Figure 2 2 sh...

Страница 12: ...Installation 2 3 7 Figure Tube support and end seal 2 3 8 Figure Tube support and end seal assembly For extra stability of the tube repeat the assembly instructions above as shown in figure 2 3 82 3 9...

Страница 13: ...ounting frame It is the customer s responsibility to carry out a risk assessment on the siting and mounting of the furnace and to ensure that it is operated in a safe manner see section 2 2 for the re...

Страница 14: ...Alignment of radiation shields is similar to that of insulation plugs see below If a metal work tube is being used in the furnace ensure that it is earthed See the safety warning in section 4 4 2 4 2...

Страница 15: ...ernal supply fuses in which case customer fusing is preferred but not essential See section 9 0 for furnace rating power current and fusing information The external connection should be either a perma...

Страница 16: ...3 0 Temperature Controller 3 0 Temperature Controller If this product is fitted with a temperature controller instructions are provided separately 16...

Страница 17: ...ct will start to heat up The Heat light s glow steadily at first and then flash as the product approaches the desired temperature or a program setpoint Over Temperature option only If the over tempera...

Страница 18: ...item 2 in Warning Labels when the furnace is being loaded or unloaded The work tube can become electrically conductive at high temperatures If an element has failed and collapsed onto the work tube t...

Страница 19: ...The controller can be set to limit both the heating and cooling rate 4 6 Pressure Work tubes are not able to accept high internal pressure When gas seals or similar fittings are in use the gas pressur...

Страница 20: ...etermine whether full calibration is required Carbolite Gero can supply these items Depending on the controller fitted the controller instructions may contain calibration instructions 5 3 After Sales...

Страница 21: ...red during maintenance or repair Whilst there is no evidence of any long term health hazards it is strongly recommended that safety precautions are taken whenever the materials are handled Exposure to...

Страница 22: ...nnections Thermocouple cable colour codings are thermocouple leg colour positive type N pink negative white Disconnect the thermocouple from its terminal block Carefully withdraw the thermocouple from...

Страница 23: ...may not be fitted If any fuse has failed it is advisable for an electrician to check the internal circuits Replace any failed fuses with the correct type For safety reasons do not fit larger capacity...

Страница 24: ...a low temperature The door switch es if fitted may be faulty or need adjustment The contactor relay if fitted may be faulty The heater switch if fitted may be faulty or need adjustment The SSR could b...

Страница 25: ...shows a low temperature The thermocouple may be faulty or may have been removed out of the heating chamber The thermocouple may be connected the wrong way around The controller may be faulty 2 Produc...

Страница 26: ...safety switch es Key F1 F2 F3 Fuses FIL Filter R1 1 R1 2 Relay Contactor R1 Relay C Temperature Controller TC Control Thermocouple SSR Solid State Relay SSW Safety Switch H Heat Lamp EL Element s SW...

Страница 27: ...Filter R1 1 R1 2 Relay Contactor R1 Relay C Temperature Controller TC Control Thermocouple SSR Solid State Relay SSW A SSW B Safety Switch es H Heat Lamp EL Element s SW Instrument Switch es N Neutral...

Страница 28: ...on EMC filter circuit board s F2 Auxiliary Circuit Fuses Fitted on board to some types of EMC filter May be omitted up to 25 Amp phase supply rating 2 Amps glass type F On board 20 mm x 5 mm Other 32...

Страница 29: ...vered by this manual are fitted with elements designed for use over the range of voltages 200 V 240 V the power limit parameter is set according to the table below These models may be relocated to a d...

Страница 30: ...le from Carbolite Gero for work tubes with outside diameters of 32 mm 46 mm and 60 mm 10 1 Environment The furnaces contain electrical parts and should be stored and used in indoor conditions as follo...

Страница 31: ...ation procedure see section 2 1 Step 3 After reading the controller operation instructions heat the furnace following the temperature profile given below This will need to be done manually on furnaces...

Страница 32: ......

Страница 33: ...Notes Service Record Engineer Name Date Record of Work...

Страница 34: ...us at the address below or ask your nearest stockist For preventive maintenance repair and calibration of all furnace and oven products please contact Carbolite Gero Service Telephone 44 0 1433 624242...