CHAMBER VOLTAGE
PLEASE WAIT
Figure 8-20 Chamber Voltage Please Wait Screen
When a measurement is available, Figure 8-21 Chamber Voltage Results Screen will
appear.
CHAMBER VOLTA
151.2V
GE
OK
ENTER to Accept
Figure 8-21 Chamber Voltage Results Screen
The measurement is compared with the value input at the factory. If the results are out of
range, the message “FAIL SEE MANUAL” appears. If this occurs, contact Capintec’s only
Authorized Service Center at 1-800-227-6832.
Press
ENTER
to accept the result. Figure 8-5 Tests Menu will re-appear.
ACCURACY TEST
To perform an Accuracy Test, press
TEST
from the Measurement Screen. Figure 8-5 Tests
Menu will appear. (Accuracy Test is also part of the Daily Test function. Reference this
chapter, SECTION: DAILY TEST: Accuracy and Constancy Test in Daily Test on page 8-7).
This will allow you to perform the test using any of the Test Sources.
Select
ACCURACY
. Figure 8-22 Accuracy Test Measure Screen will appear for each Test
Source (Co57 is used in the example).
8 - 10
BACKGROUND AND TESTS
October 13
Содержание CRC-25R
Страница 2: ......
Страница 12: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 26: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 58: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 104: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 122: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 192: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 208: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 226: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 227: ......
Страница 228: ......