CAPINTEC, INC.
CAPTUS
®
600
2. WORK AREAS
3. UNRESTRICTED AREAS
4. RECEIVING
5. SEALED SOURCES
6. GENERAL
1. BACKGROUND
7. PRINT ALL
0. RETURN TO WIPE SETUP MENU
14:32
TRIGGER LEVEL/
COUNT TIME
Figure 12-7
If
BACKGROUND
is selected, the trigger level is in cpm (cps if in Becquerel mode). If
counting rate is greater than the trigger level during a background measurement, the
message “BACKGROUND HIGH” will appear.
If
WORK AREAS
or
UNRESTRICTED AREAS
is selected, the trigger level is in dpm (Bq if in
Becquerel mode).
If
RECEIVING
or
SEALED SOURCES
is selected, the trigger level is in nCi (Bq if in
Becquerel mode).
The
GENERAL
category is for wipes in an area with no description. Only counting time is
entered; there is no trigger level for this category.
Select the appropriate area by pressing the number key that corresponds to that area. A
screen similar to Figure 12-8 will appear. (The Work Areas category is given as an example)
1 3 :0 6
W O R K A R E A S
T R IG G E R L E V E L : > 2 0 0 0 d p m
C O U N T IN G T IM E : 6 0 S E C
O K
Figure 12-8
Any peak found by the wipe test will be labeled “HIGH” if the Count Rate associated with the
Region of Interest of the peak is greater than the trigger level. Notice that the trigger level is
November 07
WIPE TEST SETUP
12-5
Содержание CAPTUS 600
Страница 2: ......
Страница 40: ......
Страница 62: ......
Страница 77: ...CAPINTEC INC CAPTUS 600 Page 1 November 07 ACCEPTANCE QUALITY ASSURANCE TESTS 6 15...
Страница 78: ...CAPINTEC INC CAPTUS 600 Page 2 Page 3 6 16 ACCEPTANCE QUALITY ASSURANCE TESTS November 07...
Страница 90: ......
Страница 102: ......
Страница 112: ...CAPINTEC INC CAPTUS 600 Figure 9 15 9 10 THYROID UPTAKE ONE CAPSULE METHOD November 07...
Страница 120: ...CAPINTEC INC CAPTUS 600 Figure 10 11 10 8 THYROID UPTAKE TWO CAPSULES METHOD November 07...
Страница 131: ...CAPINTEC INC CAPTUS 600 Figure 11 16 November 07 THYROID UPTAKE STANDARD CAPSULE METHOD 11 11...
Страница 132: ......
Страница 150: ...CAPINTEC INC CAPTUS 600 Figure 13 11 13 8 WIPE TEST November 07...
Страница 152: ......
Страница 162: ...CAPINTEC INC CAPTUS 600 Figure 15 5 15 4 LAB TESTS URINE TEST November 07...
Страница 167: ...CAPINTEC INC CAPTUS 600 Figure 16 5 November 07 LAB TESTS BLOOD TESTS 16 5...
Страница 171: ...CAPINTEC INC CAPTUS 600 Figure 16 9 November 07 LAB TESTS BLOOD TESTS 16 9...
Страница 175: ...CAPINTEC INC CAPTUS 600 Figure 16 14 November 07 LAB TESTS BLOOD TESTS 16 13...
Страница 176: ......
Страница 186: ......
Страница 208: ......
Страница 214: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......