Locate the endoscope port and follow the instructions for attaching the  

corresponding Endoscope Connector. All connectors must be attached.

Air/Water Connector

Align the connector with 

the air/water bottle port and 

push to fully engage.

2-8-650HAN/2-8-650CAS 

HOOKUP CONNECTION GUIDE

Using MEDIVATORS

®

 ADVANTAGE™ Automated Endoscope Reprocessors

Leak Test Connector

Align the slot on connector with the 

pin on the scope. Push on and turn 

clockwise to engage.  To remove 

leak test connector, turn connector 

counter-clockwise until slot is 

aligned with pin and pull off. Do not 

pull until slot is aligned with pin.

READ ALL INSTRUCTIONS PRIOR TO CONNECTION

Port ID

Port Name

Connector Assembly Description

Order Number

1

LEAK TEST

Leak Test Connector Assembly

H28650-1

2

JET/AUX

Air/Water Connector Assembly

H28650-26

3

BIOPSY

Biopsy Inlet Connector Assembly

H28650-3

4

AIR

Air Probe Connector Assembly

H28650-4

5

ELEVATOR

Forward Water Jet Connector Assembly

H28650-5

6

WATER

Air/Water Connector Assembly

H28650-26

7

SUCTION

Suction Connector Assembly

H28650-7

A

NA

Air/Water, Suction Channel Plug

N/A

If a Connector Assembly needs replacement, locate the port number/name and use 
the corresponding order number found in the Connector Assembly Identification Chart. 
Connector Assemblies are specific to individual hookups. Check each Hookup Connection 
Guide for relevant information.

Each Advantage Hookup is comprised of multiple ports and connector assemblies. Each 
port is labeled with a Port ID number (circled). This same number also identifies the 
corresponding Connector Assembly. Each Connector Assembly is fitted with a specific 
endoscope connector. (see Fig. 2).

EXAMPLE CONNECTOR ASSEMBLY 

(Figure 2)

CONNECTOR ASSEMBLY IDENTIFICATION CHART

Ferrule

Endoscope

Connector

example

Tubing

Connector 

Coupling

Label

ADVANTAGE HOOKUP AND CHANNEL SEPARATOR 

(Figure 1)



Biopsy Inlet Connector

Align the biopsy connector 

disc over the biopsy inlet 

and push down to compress 

spring and slide the frame 

forward to fully engage.

Air Probe Connector

Align with air probe, push 

and slide until fully engaged.

Suction Connector

Align silicone tubing with suction 

barb and push tubing onto barb 

until fully engaged.

Air/Water, Suction 

Channel Plug

Align and attach Peer 

Medical channel plug 

(supplied with endoscope) 

to air/water and suction 

ports. Insert and close 

clamp to secure. Ensure 

that luer caps are fitted 

and secure.

A

Forward Water Jet Connector

Align stainless steel fitting with forward 

water jet port, push and turn clockwise 

to connect.

LO

T

XXXXXX

  

ID: 

xxxX

P/N: X-X-XXXHAN



Ensure proper alignment of all connectors with the 

endoscope mating parts prior to attachment.

If you have any questions regarding this hookup, in the U.S., please contact MEDIVATORS Technical Support at  

1-800-444-4729. Outside the U.S., please contact your MEDIVATORS Distributor.

DESIGNATED HOOKUP FOR THE FOLLOWING 

ENDOCHOICE

® 

ENDOSCOPES*:

C38s

C38s-133

C38s-150

CS-1001

F1C-133

F1C-150

FUSE-1C

FUSE-1G

FUSE-F1C

FUSE-G32

* This list of corresponding endoscope models is subject to change without notice. For a full list, please go to:

www.medivators.com/hookuplookup

© 2016 Medivators Inc. 50098-562/B

Representative  endoscope  for  illustrative  purposes  only.  For  exact         

.

port locations, refer to endoscope manufacturer’s instruction guide.

Forward Water Jet Cleaning

Adapter Gas 2426

This is an optional piece

Air/Water, Suction 

Channel Plug Gas 2393

ADDITIONAL REQUIRED ENDOSCOPE CONNECTIONS