4.3 Movement Short-cut Zones
A short-cut movement is defined as, when the OTC moves directly to the intended position,
without first moving up into the transport interval zone.
Movement short-cut zones are defined both for the Table and the Wallstand, as rectangular
cubes.
To perform a short-cut movement, the following conditions must be fulfilled;
• The System shall be positioned above the Table, or at the specified short-cut zone in front
of the Wallstand.
• At least one of the OTC corners must be inside the short-cut zone at the start of the
movement.
Outside these short-cut zones, the OTC will first move in Z direction, to the transport inter-
val zone, before moving to position.
Содержание OMNERA 400A
Страница 1: ...Technical Publication Operator s Manual 0073 095 418 EN Rev 5 1 ...
Страница 14: ...OMNERA 0073 095 418 EN Rev 5 1 Operator s Manual Page 8 2020 09 25 Introduction System Description ...
Страница 40: ...OMNERA 0073 095 418 EN Rev 5 1 Operator s Manual Page 34 2020 09 25 Safety Electromagnetic Compatibility EMC ...
Страница 84: ...OMNERA 0073 095 418 EN Rev 5 1 Operator s Manual Page 78 2020 09 25 User Interface Wall stand Control Elements ...
Страница 107: ...Operator s Manual 0073 095 418 EN Rev 5 1 OMNERA 2020 09 25 Page 101 Operating the System Detector Wallstand ...
Страница 118: ...OMNERA 0073 095 418 EN Rev 5 1 Operator s Manual Page 112 2020 09 25 Operating the System Portable detector table ...
Страница 130: ...OMNERA 0073 095 418 EN Rev 5 1 Operator s Manual Page 124 2020 09 25 Function and Safety Checks Monthly Checks ...
Страница 138: ...OMNERA 0073 095 418 EN Rev 5 1 Operator s Manual Page 132 2020 09 25 Technical Specifications Cabinet ...
Страница 140: ...OMNERA 0073 095 418 EN Rev 5 1 Operator s Manual Page 134 2020 09 25 Waste Disposal ...
Страница 150: ...OMNERA 0073 095 418 EN Rev 5 1 Operator s Manual Page 144 2020 09 25 Appendix A Glossary ...
Страница 156: ......