Guidance and manufacturers declaration - electromagnetic immunity
Radiated RF
IEC 61000-4-
3
3 V/m
80 MHz to 2,5 GHz
3 V/m
80 MHz to
2,5 GHz
d = 1,2 √p
80 MHz to 800 MHz
d = 2,3 √p
800 MHz to 2,5 GHz
where
p
is the maximum output power rating
of the transmitter in watts (W) according to the
transmitter manufacturer and
d
is the recom-
mended separation distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
a
should be range.
b
Interference may occur in the vicinity of equip-
ment marked with the following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be pre-
dicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF trans-
mitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the System is used exceeds the applicable RF compliance level above, the System
should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as relocating the System.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
Recommended separation distances between portable and mobile RF communications equip-
ment and System
The System is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the System can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the System as recommended below, according to the maximum output power of
the communications equipment.
Rated maximum output power
of transmitter W
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d = 1,17 √p
80 MHz to 800 MHz
d = 0,35 √p
800 MHz to 2,5 GHz
d = 0,7 √p
0,01
0,12
0,04
0,07
0,1
0,37
0,11
0,22
1
1,17
0,35
0,7
10
3,69
1,11
2,21
Содержание OMNERA 400A
Страница 1: ...Technical Publication Operator s Manual 0073 095 418 EN Rev 5 1 ...
Страница 14: ...OMNERA 0073 095 418 EN Rev 5 1 Operator s Manual Page 8 2020 09 25 Introduction System Description ...
Страница 40: ...OMNERA 0073 095 418 EN Rev 5 1 Operator s Manual Page 34 2020 09 25 Safety Electromagnetic Compatibility EMC ...
Страница 84: ...OMNERA 0073 095 418 EN Rev 5 1 Operator s Manual Page 78 2020 09 25 User Interface Wall stand Control Elements ...
Страница 107: ...Operator s Manual 0073 095 418 EN Rev 5 1 OMNERA 2020 09 25 Page 101 Operating the System Detector Wallstand ...
Страница 118: ...OMNERA 0073 095 418 EN Rev 5 1 Operator s Manual Page 112 2020 09 25 Operating the System Portable detector table ...
Страница 130: ...OMNERA 0073 095 418 EN Rev 5 1 Operator s Manual Page 124 2020 09 25 Function and Safety Checks Monthly Checks ...
Страница 138: ...OMNERA 0073 095 418 EN Rev 5 1 Operator s Manual Page 132 2020 09 25 Technical Specifications Cabinet ...
Страница 140: ...OMNERA 0073 095 418 EN Rev 5 1 Operator s Manual Page 134 2020 09 25 Waste Disposal ...
Страница 150: ...OMNERA 0073 095 418 EN Rev 5 1 Operator s Manual Page 144 2020 09 25 Appendix A Glossary ...
Страница 156: ......