Alignment patient, LLS-1 and the X-ray beam
•
Align the X-ray center beam on the
vertical centerline
in the middle of the
stand (left-right)
•
Position the patient in front of the LLS-1
stand.
•
Adjust the LLS-1 stand
horizontal
centerline
in the middle of the target
area. Use the up-down buttons from
hand control to adjust the proper height
of the stand.
•
Align the X-ray center beam on the
horizontal centerline in the middle of the
stand (move the X-ray tube up or down).
•
Collimate the target area.
Note:
To prevent overlap artefacts, stay in
between the center area with the X-ray
center beam.
•
FFD > 2.0m, depending of the used
collimator. Normally this is around
~2.5m
Note:
When using only 2 detectors:
•
Align the X-ray center beam on
horizontal line in the middle of the
two used detectors. (Upper or lower
Centerline)
Note:
To prevent overlap artefacts, stay in between the center area with the X-ray
beam center.