20.3.2  Disinfecting the system 

This subsection describes the procedures and precautions for disinfecting the system. 

For disinfection of transducers, refer to the operation manuals provided with the 
transducers. 

DANGER:

   After disinfecting the system, ventilate the room fully before 

turning the system ON. 
If a flammable or explosive gas remains near or in the system, 
it may cause a fire or explosion.

 

WARNING

:   1.  Before disinfecting the system, be sure to disconnect the 

system power cable plug from the outlet of the facility. 
If the system is defective, an electric shock may occur. 

 

2.  Do not spill or spray liquids such as water onto the 

system or peripheral units.  If a liquid such as water 
enters the system or peripheral units, an electric shock 
may occur. 

 

CAUTION:

   1.  Observe the following precautions to prevent infection. 

 

  Wear sterile protective gloves. 

 

  Wear new sterile protective gloves each time disinfection 

is performed. 

 

2.  Do not disinfect the system using a method or chemical 

other than those specified in this manual.  Doing so may 
result in the system not being disinfected properly or in the 
surface of the system deteriorating or becoming discolored. 

 

3.  After chemical disinfection, dry the surface of the system 

fully. 

 

4.  The efficacy of disinfectants is not guaranteed by Canon 

Medical Systems.  Contact the manufacturer of the relevant 
chemical for information on its efficacy. 

 

5.  Determine the effectiveness of the chemical based on the 

criteria (effective period, number of times of use, 
discoloration of the chemical, results of testing using 
effectiveness test kits, etc.) described in the documentation 
provided by the manufacturer of the chemical. 

 

NOTICE:

 

1.  When chemical disinfection is repeated over time, the surface of the 

system changes gradually. 

2. If any abnormality related to functions of the product is observed after 

disinfection, stop using the product and contact your Canon Medical 
Systems service representative for repair. 

Содержание Aplio i600

Страница 1: ...g the equipment After reading keep this manual in an easily accessible place CANON MEDICAL SYSTEMS CORPORATION 2017 2021 ALL RIGHTS RESERVED Issued 2021 01 OPERATION MANUAL FOR DIAGNOSTIC ULTRASOUND SYSTEM MODEL TUS AI600 FUNDAMENTALS 2B771 191EN Q For cleaning and disinfection only the following maintenance items are extracted 20 3 Preventive Maintenance Performed by the User ...

Страница 2: ...r than those specified to clean the system If this is not observed deterioration or discoloration of the surface of the system may result NOTICE Do not use solvents such as paint thinner benzine or alcohol or abrasive cleansers to clean the system These substances may cause deterioration or discoloration of the system Cleaning the transducer Clean disinfect and sterilize the transducer referring t...

Страница 3: ...pe the connector with a dry cloth only Clean the reference signal cable regularly a Wipe off all organic materials such as blood or other body fluids from the reference signal cable using a soft cloth moistened with any of the following as appropriate Purified water Mild detergent diluted with purified water Ethanol for disinfection 76 9 to 81 4 vol at 15 C b Wipe the cables with a clean soft dry ...

Страница 4: ...oth and then dry them Do not use heat to dry the electrodes c Confirm that the ECG limb clips show no signs of damage deformation or peeling Cleaning the PCG sensor and the pulse sensor NOTICE Do not immerse the PCG sensor or the pulse sensor in water or chemical solution If moisture enters the connector the system may fail PCG sensor Pulse sensor Clean the PCG sensor and the pulse sensor regularl...

Страница 5: ...lure a Monitor cover Clean the monitor cover using a soft cloth moistened with mild detergent b LCD screen Use a soft cotton cloth or lens cleaning paper to clean the LCD screen If stains on the LCD screen are difficult to remove wipe the screen gently with a soft cloth slightly moistened with water or mild detergent and then wipe it dry with a soft dry cloth Cleaning the surface of the touch pane...

Страница 6: ...ain panel using cotton buds c Wipe the transducer connector with a soft dry cloth If it is difficult to remove all the stains with a dry cloth use a soft cloth moistened with water To clean the trackball a Turn the trackball ring in the direction of the arrow in the left figure below and remove it together with the trackball NOTICE To prevent malfunction be careful when removing the ring ...

Страница 7: ...dvertently c Remove the inner ring dustproofing from the ring and remove any dust and dirt on the contact surface d Remove all foreign matter from the ball and from the interior of the case using a soft cloth or gauze Do not use sharp or hard materials to clean inside the case In addition use a microfiber cloth to clean lens cover sections left and bottom ...

Страница 8: ...191EN Q 232 e Set the ball in the ring NOTE Be sure to insert the inner ring when setting the ball If the inner ring is not fitted properly the ball may not rotate properly Be sure to fit the inner ring properly ...

Страница 9: ...ction of the arrows until it is locked fixed h After the trackball is installed check the following Operate the ball and confirm that there is no unevenness in the operating force required that there is no noise during operation and that the ball does not come loose during operation Turn ON the power and confirm that the trackball operation is normal ...

Страница 10: ...reduction in system performance reliability due to clogging of the filter NOTICE Do not boot up the system while the air filter is removed The system may malfunction a Remove the air filter 1 location b Remove all dust from the filter using a vacuum cleaner at a location away from the system c Remount the air filter ...

Страница 11: ...w sterile protective gloves each time disinfection is performed 2 Do not disinfect the system using a method or chemical other than those specified in this manual Doing so may result in the system not being disinfected properly or in the surface of the system deteriorating or becoming discolored 3 After chemical disinfection dry the surface of the system fully 4 The efficacy of disinfectants is no...

Страница 12: ...are Bleach Germicidal Cleaner Effervescent chlorine tablets sodium dichloroisocyanurate NaDCC BIOSPOT 5000 ppm Not available in the USA Isopropyl alcohol 70 vol Ethanol for disinfection 76 9 to 81 4 vol at 15 C Reference signal cable and ECG limb clips clip type electrodes Alkyldiaminoethylglycine hydrochloride 0 5 Benzalkonium chloride 0 2 w v Benzethonium chloride solution 0 2 w v Chlorhexidine ...

Страница 13: ...ctant For detailed information concerning handling of the disinfectant refer to the manual provided with the disinfectant To ensure that the disinfectant is effective confirm the working concentration replacement interval and temperature of the disinfectant To ensure that the disinfectant is effective be sure to observe the use criteria such as the effective period number of times to be used disco...

Отзывы: