Update CAV 4.2 Software and Firmware
Software and firmware code updates may occasionally need to be manually installed on the CAV 4.2
user-interface, controller boards, and controlling PCs. There are four types of code that can be installed:
•
UI updater: this software is actually the complete UI updater program (UpdateUI.exe) that runs on
the CAV 4.2. It is used to install the UI software and firmware updates. It is updated by copying it
over the existing UpdateUI.exe.
•
CAV 4.2 GUI software: this software is the CAV 4.2 user-interface. To update the CAV 4.2 GUI
software, you must be logged out of the user-interface and running the UI updater instead.
•
Firmware updates: this code controls the baths, sample handlers, and valve controls. It is installed
directly onto the respective controller boards using the UI updater.
•
VISCPRO II: this software is installed on the controlling PC.
Tools and Materials Required
See Table 28 for tools and materials required to update the software and firmware.
Table 28 — Tools and Materials Required for Software and Firmware Update
Setting
Description
#2 Philips screwdriver
USB flash drive
Software/firmware updates
USB keyboard
USB Type A-Male to Mini B cable (6ft)
Prepare the USB Flash Drive
A USB flash drive must be loaded with the software and firmware to be updated.
1.
On a PC, download the update packages from the CANNON website Software & Downloads page.
2.
Unpack and copy the update files to the USB flash drive. Only one set of files can be on the USB flash
drive during an update.
Replace the UI Updater
1.
From the GUI, click
Help > Upgrade UpdateUI
.
2.
Plug the prepared USB flash drive into the USB connection.
3.
Follow the instructions on the screen.