CANNON Instrument Company® | Operation
41
View Test Results (VISCPRO II)
Data obtained during sample testing is temporarily displayed in the
Sample Action
field and
permanently stored in the data base, which may be viewed by opening an analysis package.
Tests must finish running before they can be viewed. The entire tray does not need to have finished, but
individual results are only available in the database after they have completely finished running.
1.
In the
VISCPRO II Menu
, select
Analyses > View Analysis
. The
Choose Analysis
window opens.
2.
Open
Basic Package
and select
CAV Data Table
.
3.
Select the
Open Select Analysis Configuration
radio button.
4.
Select an analysis and click
OK
.
Control Preheaters
The preheaters are associated with the sample test. Preheater settings can be modified by editing the
configuration of a specific sample test or editing the test method template. See Table 21.
Table 21 — Preheater Settings
Parameter name
Description
Preheat Type
No preheat
Minimum preheat
starts the preheat cycle as soon as the
previous sample begins running and runs for at least the
set preheat time.
Serialized preheat
starts the preheat cycle after the
previous sample finishes and runs for exactly the set
preheat time.
Minimum Preheat Time
This is the time that the preheater is applied to the
sample.
Preheat Temperature Units
Select either Fahrenheit or Celsius.
Preheat Temperature
Fahrenheit: 0 °F to 437 °F; Celsius: 0 °C to 225 °C
The heater temperature is determined by the
Preheat
Temperature
setting of the sample that is in the preheat
position, even if that sample is set to
No preheat
. Set the
temperature to 0 (zero) to turn off the heater.