Priming possible: only at strat of a new infusion or start of a new partial infusion.
• Access to PRIME phase.
• PRIME
(each depressing administers 0.5 ml - max 1.5 ml).
• PRIME phase finish, infusion start.
• PRIME interruption (if in execution).
• Device switch OFF (if PRIME carried out).
In case there is a need to prime one or more times during an
already started infusion, the clinician bolus can be used
PRIme
10
20
10
20
10
20
10
20
10
20
10
20
PUmP ACTIvATION
BUTTONS
• The screen will show all symbols.
• Piston self-adjust.
• Pump switches OFF.
10
20
10
20
10
20
ReveRSINg Of THe PUmP PISTON
NUmBeR Of INfUSIONS DelIveReD
P
U
m
P
O
f
f
• Reversing of the pump piston to the start position.
• Reading the number of infusions delivered.
• Reset the number of bolus delivered (L0).
Confirm the reset of bolus delivered.
• Pump switches ON.
• Press for 7 seconds.
• Key-pad lock level selection.
• Selection of the lock level.
and
and
depress simultaneously
depress simultaneously
depressed for 8 seconds
press without releasing
button
PUmP ON
keY-PAD lOCk-OUT
10
20
10
20
10
20
10
20
10
20
10
20
10
20
10
20
10
20
SeTTINgS
P
U
m
P
O
f
f
P
R
Im
e
C
O
N
D
IT
IO
N
Operations feasible only with key-pad unlocked.
•
Setting flow
(from 0.05 to 35 ml/h with the following
increments:- from 0.05 to 10 with
0.02
ml increments; - from 10
to 35 with 0.05 ml increments).
•
Setting bolus dose
(from 0.10 to
9.98
ml with
0.02
ml
increments).
•
Setting bolus time limitations
(- from 5’ to 1 h with
increments of 5’ - from 1 h to 24 h with increments of 15’).
•
Setting boluses number limitations
(0-12 per h).
Bolus lock-out possibilities: combination of bolus time and
bolus number limitations. The computer always chooses the
programming that results in the smallest bolus volume.
•
Setting partial volume
(from 1 to 50 ml with
increments of 1 ml - possible only at the begin of a new
infusion).
•
Decrease/Increase above parameters.
1st depressing
2nd depressing
3rd depressing
4th depressing
5th depressing
/
and
and
depress simultaneously
depress simultaneously
SCReeN
BUTTONS
SCReeN
BUTTONS
SCReeN
BUTTONS
SCReeN
IN
S
e
R
T
IN
g
B
A
T
T
e
R
Y
P
H
A
S
e
PHYSICIAN SHORT INSTRUCTION
/