background image

Candela Vbeam Perfecta, Platinum, Aesthetica Treatment Guidelines 

 

8502-00-0891 Revision G            Candela Proprietary                             Sheet 1 of 8 

 

 

 

Vbeam Perfecta, Platinum, Aesthetica Treatment Guidelines 

These guidelines were developed from clinical experience and are subject to change as additional experience 
is gained. Be sure to inquire with your Candela Sales Representative, Clinical Consultant or visit 
MyCandela.com regularly for the latest updates and a comprehensive bibliography list of references/published 
articles. Treatment Guidelines do not take the place of the procedures and instructions found in the Operator’s 
Manual. FAILURE TO USE THE LASER IN ACCORDANCE WITH SUCH PROCEDURES AND 
INSTRUCTIONS COULD RESULT IN SERIOUS INJURY TO THE OPERATOR, THE PATIENT AND 
OTHERS, AS WELL AS DAMAGE TO THE LASER SYSTEM. Follow OSHA and ANSI standards for laser 
safety. Protective eyewear must be worn by all persons in the treatment room during laser operation. Perform 
User Verification tests as outlined in the Operators Manual at the start of each treatment day and when the 
hand piece is changed. Check the delivery system for any damage (i.e.: dropped)

. Discontinue use of your 

laser delivery system if you suspect a problem. 
     Lesion 

Spot 

size 

(mm) 

Pulse 

Duration 

(ms) 

Fluence 

(J/cm

2

DCD 

Spray

 / 

Delay

 

(ms) 

Retreat 

(weeks) 

Comments  

Telangiectasia 

 
Facial 
 
 
 
 
 
 
Nasal 
 

 
 
 
 
 
 
 
 

 


10 

3x10 

 
 
 

5, 7 

10 

3x10  

 
 
 
 
 
 

 

 

 

6,

10

,20 

6,

10

,20 

6,

10

,20 

10 

 
 
 

40 

 
 

 
 
 
 
 
 
 

 

7.5-12 

7-12 

5.5-7.5 

11-14 

 
 
 

8-12 

6.5-7.5 

10-12 

 
 
 
 
 
 
 

 

 medium 

30/20 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 
 
 
 
 

4-6 

 
 
 
 
 
 
 
 
 
 
 
 

Double or triple pulse 
depending on the refill of 
the vessel. 
Use 3x10 on linear vessels. 
For vessels which do not 
clear immediately: treat first 
with the round spot followed 
by the 3x10 spot. 
Use of shorter pulse 
durations will induce 
purpura but may clear 
vessels faster. 
 

Leg  
 
 
 

 

 
 

3x10 

 
 

 

10

,20,30, 

 

---------- 

 

40

 

11

-18 

 

------------ 

 

15-25

 

   medium 
   30/20       
------------- 

high 

40/20 

 
 
 

8-12 

 
 

 
Select longer pulse 
durations to decrease 
purpura. 

Telangectatic 
Matting 

 

10 

 
 

10

,20,30 

 

 

 

11

-15 

 

6-

7

 

 
 

medium  

30/20 

6-10 

Larger spot size may 
improve results. 
Expect erythema. 

Содержание Vbeam Perfecta

Страница 1: ...be worn by all persons in the treatment room during laser operation Perform User Verification tests as outlined in the Operators Manual at the start of each treatment day and when the hand piece is ch...

Страница 2: ...ear and become bluish pink dusky erythema May use 10 12mm spot then retreat larger vessels immediately with smaller spot size Press the lens firmly onto the skin until the skin turns pale Clinical end...

Страница 3: ...fluences and DCD for neonates and infants Clinical endpoint purpura Use lower fluences on neck and extremities Ectatic vessels associated with hypertrophic PWS may be double pulsed Warts Non Facial F...

Страница 4: ...DCD spray should be sufficient to cover the treatment area and provide epidermal cooling in opposition to the heat generated from the laser It is important to maintain both skin protection and patien...

Страница 5: ...0920 23 0814 e Seizure disorders Do not treat patients with a history of light triggered seizures f Medications Daily anticoagulation therapy iron supplements herbal supplements such as ginkgo ginseng...

Страница 6: ...cted treatment response is noted B Procedure for establishing the Purpura Free Threshold 1 Select spot size pulse duration fluence and DCD settings 2 Apply escalating fluences until purpura appears th...

Страница 7: ...e used to draw a treatment area grid When treating anal area place wet gauze over the anus first Methane gas is flammable 8 The Cleaning and Disinfecting Policy is outlined in Section 6 of the Operato...

Страница 8: ...in subsequent crusting blistering or other adverse reactions Crescents may result from not holding the distance gauge properly If problem persists discontinue treatment and refer to user verification...