New GentleYAG
Installation Procedure
Candela Corporation
8503-01-0831,Revision B
Candela Corporation Proprietary
Page2 of 6
WARNING
FOLLOW THE INSTALLATION STEPS IN SEQUENCE TO ELIMINATE ANY
POSSIBLE DAMAGE TO THE SYSTEM.
Equipment Needed:
•
Adjustable Wrench or 9/16 Socket Wrench
•
2.5 Liters of
Distilled
Water
1. Laser System Unpacking:
1.1. Using the wrench, remove the front of the crate. Keep the front nearby to use
as a ramp to roll the laser system from the crate.
1.2. Remove the accessories located on each side of the laser system from the
crate.
1.3. Place one end of the front of the crate on the floor. Butt the other end against
the floor of the crate. Grasp each handle and gently roll the laser out of the
crate.
WARNING:
Rear wheels must be
locked when removing laser from
crate. Failure to have rear wheels
locked could result in the laser
coming off the ramp.
1.4 Move the laser to the room that has been prepared for its use. Make sure the
required space and electrical requirements are met. To prevent the laser
system from moving inadvertently, lock each front wheel.