
The 9660’s overload recovery time is approximately 40
µ
s with the Rise time set to
5.6
µ
s and Flat Top set to 0.8
µ
s and using a Canberra Model 2101 preamp and
60
Co
source.
User’s Manual - ICN 9231014G
75
Operation with Reset Preamps and Selecting Inhibit Mode
Figure 36 Setting Reset Preamp Inhibit Pulse
Width
Scope:
Horizontal: 20
µ
s/div
Vertical: 50 mV/div
Vertical: 5 V/div