3. Operating Instructions

The Model 3102D is factory set for positive output polarity. If negative polarity is re-
quired, change the internal polarity board before installing the unit in the NIM bin (re-
fer to the Internal Controls section, below, for instructions).

Installation

The Canberra Model 2000 NIM bin and Power Supply or equivalent power supply
conforming to DOE/ER-00457T will accommodate the Model 3102D. The right side
cover of the unit acts as a guide for insertion of the unit. Secure the module in place by
turning the two front panel captive screws clockwise until finger tight. It is recom-
mended that the NIM bin power switch be OFF whenever any module is installed or
removed.

The Model 3102D can be safely operated where the ambient air temperature is
between 0

°

C and +50

°

C. Perforations in the top and bottom of the unit’s frame

permit cooling air to circulate through the module. When rack mounted along with
other heat generating equipment, be sure to provide adequate clearance for a flow of
cooling air through the NIM bin.

Internal Controls

Remove the unit’s left side cover to change the internal controls. Figure shows the
position of the internal controls.

WARNING

The Model 3102D generates lethal high voltage, which will
be present for several minutes after the unit is switched
off.

Before removing the unit from the NIM bin, set the VOLTAGE control to 0.00 volts
and the HV ON/OFF switch to OFF.

Avoid the risk of injury. Leave the unit in the NIM bin until the voltage reduces to
0 volts, as indicated on the front panel digital display.

5

Installation

Содержание 3102D

Страница 1: ...Model 3102D HV Power Supply User s Manual 9231570A ...

Страница 2: ...ing as part of a submission to regulatory or legal authorities pursuant to the purchaser s legitimate business needs No material in this document may be copied by any third party or used for any commercial purpose or for any use other than that granted to the purchaser without the written permission of Can berra Industries Inc Canberra Industries 800 Research Parkway Meriden CT 06450 Tel 203 238 2...

Страница 3: ...ection 6 J4 Brightness Control 6 Inhibit Control 6 Automatic Overload Shutdown 7 Setup 7 Divide By 10 Output 7 Preventive Maintenance 8 4 Circuit Description 9 A Specifications 10 Inputs 10 Outputs 10 Controls 10 Indicators 10 Performance 11 Connectors 11 Power Requirements 12 Physical 12 Environmental 12 B Installation Considerations 13 ...

Страница 4: ...Notes ii ...

Страница 5: ...m 15 to 2000 V dc and the other from 1 5 to 200 V dc The output voltage is measured and displayed by a three digit voltmeter In addition this unit al lows the user to select the output voltage polarity with an internal control The 3102D can withstand any overload or short circuit for an indefinite period of time It can be set using an internal jumper to allow for a manual reset or an automatic res...

Страница 6: ...ls and Connectors This is a brief description of the front panel controls and connectors For more detailed information refer to Appendix A Specifications 2 Controls and Connectors Figure 1 Front Panel Controls ...

Страница 7: ...This is a brief description of the rear panel connectors For more detailed information refer to Appendix A Specifications 3 Figure 2 Rear Panel Controls ...

Страница 8: ...This is a brief descriptions of the internal controls For more detailed information re fer to Appendix A Specifications 4 Controls and Connectors Figure 3 Internal Controls ...

Страница 9: ...oved The Model 3102D can be safely operated where the ambient air temperature is between 0 C and 50 C Perforations in the top and bottom of the unit s frame permit cooling air to circulate through the module When rack mounted along with other heat generating equipment be sure to provide adequate clearance for a flow of cooling air through the NIM bin Internal Controls Remove the unit s left side c...

Страница 10: ...c 0 0 7 V to the rear panel INHIBIT connector The front panel INHIBIT STATUS indicator lights when the output has been inhibited Jumper J5 The INHIBIT input is compatible with all Canberra preamps for either position of Jumper plug J5 it isn t necessary to set jumper plug J5 to match the HV INHIBIT logic high level of the associated Canberra preamp Jumper J5 selects the pull up clamp voltage level...

Страница 11: ... NIM bin power to ON the appro priate polarity LED should light indicating the polarity selected Set the HV ON OFF switch to ON and the VOLTAGE control to the desired setting The front panel digital display will indicate the HV OUTPUT voltage in kilovolts kV Divide By 10 Output The 3102D has two rear panel output connectors HV OUTPUT and 10 OUTPUT Canberra recommends that only one of these outputs...

Страница 12: ...l of the unit may be cleaned Remove power from the unit before cleaning Use only a soft cloth dampened with warm water and make sure the unit is fully dry before restoring power Because of access holes in the NIM wrap DO NOT use any liquids to clean the wrap side or rear panels 8 Operating Instructions ...

Страница 13: ...high voltage dc to dc converter is obtained from a conventional NIM power supply and uses 12V dc and 24 V dc An oscillator determines a high frequency 37 kHz at which all amplification high voltage transformation rectification and filtering occurs The amplification is a func tion of a control voltage which performs the functions of control and regulation A sample of the output voltage is compared ...

Страница 14: ...panel SHV connector 10 OUTPUT 1 5 to 200 V dc continuously adjustable Zout 20 MΩ rear panel SHV connector Controls ON OFF Front panel toggle switch enables or disables output RESET Pushbutton restores normal operation following a latched Inhibit and or overload fault condition VOLTAGE Front panel 10 turn control permits continuous adjustment of the output voltage POLARITY Internal programming plug...

Страница 15: ...NT 50 ppm C after 30 minute warmup REGULATION 0 001 variation in output voltage over the load range and 0 001 for 0 1 input voltage change within the operating range at constant ambi ent temperature OVERLOAD PROTECTION Power supply will withstand any overload including a short circuit for an indefinite period CURRENT LIMIT 1 3 mA maximum DIAL ACCURACY 1 of full scale METER ACCURACY 0 6 of full sca...

Страница 16: ...ngle width module Physical SIZE Standard single width NIM module 3 43 x 22 12 cm 1 35 x 8 71 in per DOE ER 0457T NET WEIGHT 1 4 kg 3 1 lb SHIPPING WEIGHT 2 4 kg 5 3 lb Environmental OPERATING TEMPERATURE 0 to 50 C OPERATING HUMIDITY 0 80 relative non condensing Tested to the environmental conditions specified by EN 61010 Installation Category I Pollution Degree 2 12 Specifications ...

Страница 17: ...ffects of RFI and EMC on the system However care should be taken to maintain full compliance These considerations include A rack or tabletop enclosure fully closed on all sides with rear door access Single point external cable access Blank panels to cover open front panel Bin area Compliant grounding and safety precautions for any internal power distribution The use of CE compliant accessories suc...

Страница 18: ...Notes 14 Installation Considerations ...

Страница 19: ... OR IN THE CASE OF DEFECTIVE SERVICES REPERFORMANCE AT CANBERRA S COST IS YOUR SOLE AND EXCLUSIVE REMEDY UNDER THIS WARRANTY EXCLUSIONS Our warranty does not cover damage to equipment which has been altered or modified without our written permission or damage which has been caused by abuse misuse accident neglect or unusual physical or electrical stress as determined by our Service Personnel We ar...